EQAWarden · validation

EQAWarden validation evidence, laboratory ratification and change control

EQAWarden's validation evidence in one place:

  • The release it covers
  • How its results are tested
  • Where each threshold is defined, and its source
  • What the laboratory's Chemical Pathologist reviews and ratifies before suggestions are used
  • How changes are controlled afterwards
The release

Release and section versions on each assessment

EQAWarden is built in versioned sections.

  • The release number fixes the version of each one
  • Each section's files are fingerprinted
  • A copy can therefore show that it matches the release it identifies
Release 2026.40

The current release. Every copy shows it in its footer and on its Status page. It is checked at start against its own file list.

Engine 4.16

It covers the statistics, scoring and patterns. It also covers the lot and internal quality control (IQC) linkage, the interpretation and the suggested outcomes.

Knowledge 7.2

It holds what the engine reads: the rules, thresholds and cause weights. It also holds the analytes and programmes. The analytical performance specification (APS) defaults and the references are there too.

A release is set only when every section's files match their recorded fingerprints and every check below has passed.

The release check writes a validation dossier with:

  • The results
  • The fingerprints and the SHA-256 of every file
  • A sign-off block
Test cases

A fictional year and fifteen golden scenarios checked against known answers

The test data is generated, not taken from any laboratory:

  • One RCPAQAP-style General Serum Chemistry cycle of 21 surveys with two samples each
  • The cycle covers 26 analytes and 1,092 results
  • About ten months of daily internal QC (15,602 points with 98 events)
  • A lot history

The generator checks that the fixture files are exactly what it produces, so a changed fixture is detected.

The golden scenarios

Each states the findings that must appear, the cause that must rank first and the outcome that must be suggested.

  • A reagent lot step with concordant internal QC (creatinine)
  • A proportional bias with concordant Unity peer data (calcium)
  • Two samples transposed (potassium), and a whole survey of tubes swapped
  • Results in the wrong unit (glucose in mg/dL) and a decimal slip (ALT)
  • A reconstitution error: one vial 6% low across its analytes
  • A one-off result outside with stable internal QC (sodium)
  • Drift within the APS (ALT)
  • A method group of four (TSH)
  • A method-dependent analyte against the all-method median (albumin, free T4)
  • A known handling pitfall (bicarbonate losing CO2)
  • A clean year with nothing to find (magnesium)
  • A censored result that must not be scored (TSH)

What is checked

  • Each of the 1,092 scores and zones against its expected answer
  • Each statistic against reference values
  • The same input gives the same output on repeated runs
  • Each assessment recomputed with later results removed gives the same answer, so an assessment can be recomputed as at its survey from the same retained data
  • That each threshold in the code equals the value in the knowledge files
  • End to end: imports and reports
  • End to end: the review record and the audit chain
  • End to end: the security guards
  • That the test suite itself catches faults planted on purpose
Proving a change

Comparing every output before and after a change

Some changes should leave results alone. Examples are a relabelling, a tidier rule file or a fix elsewhere. Before such a change is released:

  • Each output EQAWarden gives on the fictional year is produced by the old release and the new one
  • The two are compared item by item
  • Each assessment, survey and analyte view is compared in full
  • So is each internal QC view and score

When a change is meant to alter results, each difference is listed and accounted for.

Every output

Each of the 1,092 scores is compared too. Only the version stamp is set aside.

Knowledge 5.1

The knowledge labels were corrected to say who owns each choice. Then 1,735 outputs were compared. All were identical.

Engine unchanged

The network fixes of release 2026.11 left the engine's files as they were, as the section fingerprints show.

Threshold definitions

Each threshold defined once, with its owner and its reason

Each threshold, window and weight the engine uses is defined once. Each has an owner and a reason. The owner is one of these:

  • RCPAQAP or ISO 13528
  • IUPAC or NIST
  • Westgard, Bio-Rad or Fraser
  • A dated laboratory decision
  • An EQAWarden default

The defaults:

  • 73 of the 118 parameters are EQAWarden defaults
  • Those defaults are what the laboratory ratifies
  • The sign-off pack below holds the full list with every value

Who owns each number

Where each number comes from, and how the defaults become the laboratory's. Thresholds and their owners

The rules

What each finding rule looks for, and what the reviewer sees when it applies. The rules

The arithmetic

What the statistics assess over surveys and the cycle, and what small numbers cannot show. The statistics

Laboratory sign-off

Ratification by the laboratory's Chemical Pathologist before suggestions are used

This is a step each laboratory takes for its own use of EQAWarden.

  • The sign-off pack is generated from the knowledge files of the release in use, so it matches what runs
  • The Chemical Pathologist marks each item Agree or writes the change wanted
  • A senior scientist they delegate may do this instead
  • They sign with their name, role and registration
  • The signature carries the date and the laboratory's QMS reference
  • Until it is signed, the laboratory does not use EQAWarden's suggestions in review records

EQAWarden records the sign-off on its Status page with an electronic signature. It is recorded against a fingerprint of the knowledge the pack was made from.

  • It is recorded as Ratified, or as Ratified for an earlier release
  • Otherwise it is recorded as Not yet ratified
  • Until a sign-off is recorded for the knowledge in use, the assessment and review screens say the suggestions are not yet ratified by the laboratory
  • The status marks reviews; it does not block them
  • When a later release changes items in the pack, they are listed by number for signing again
  • Each signed review records whether the suggestions were ratified when it was signed
Thresholds EQAWarden choseThe 73 default parameters. Each has its value, its reason and where it is used.
Cause weightsHow strongly each finding counts towards each cause, so the laboratory sees and can change why one cause ranks above another.
Rule choicesThe choices made where rules interact, and the decisions left open for the laboratory.
APS rows to confirmThe default APS rows that could not be verified against a current RCPAQAP publication.
Weekly QC settingsThe internal QC review's limits, if the laboratory has its own.
Textbook factsThe statements in the knowledge base resting on general textbook knowledge, to be checked against the laboratory's own reference text.
For informationThe thresholds EQAWarden does not set. They are owned by RCPAQAP, the standards or the laboratory's earlier decisions.

EQAWarden's outputs are suggestions; the laboratory decides how they are used.

Change control

Engine changes made in a new release, not in a running copy

Changed in the files

When the laboratory changes a threshold or weight of the engine, the change is made in the knowledge files. It gets a new version and is released.

  • Each release checks that the program and the files agree
  • The weekly QC limits and the APS table are changed in the program
  • So are QC targets and laboratory settings
  • Those changes are made only by an Authoriser at Manager or Chemical Pathologist level
  • Each needs a reason and an electronic signature, and is kept on the change trail

Major or minor

A section's major version moves when it can give a different result or record for the same input.

  • Its evidence is then verified again before deployment
  • The minor version covers wording, speed and fixes with results unchanged

Recorded on every review

Each assessment and each signed review record carries the release and section versions it was made under. A record can then be matched to the evidence for that release.

New releases. EQAWarden does not update itself, by design.

  • A new release is a new program folder, installed by the laboratory's IT
  • The data is carried over, and the folder checked against the release's published fingerprints
  • Licences are renewed by file

Deployment and licensing.

LotWarden

LotWarden release and validation

LotWarden 2026.68 is released the same way:

  • Versioned sections with fingerprints
  • A release check re-run before every validation dossier
  • A dossier that marks itself DRAFT if anything disagrees

Its Reader and its network mode are tested as well.

  • Its Reader is scored against answer keys before release
  • Its Reader accuracy page measures reading performance on the laboratory's own delivery notes once they are in use at the site
  • Its network mode was tested under failure: network reliability testing

LotWarden's validation and change control · Reader accuracy

Validation dossier and sign-off pack on request

Both are generated for the release in use. Your Chemical Pathologist is welcome at the demonstration.