Statistical methods and assessment criteria
EQAWarden uses established statistical methods to support external quality assessment (EQA) review. This page says in plain words what they look at, what the reviewer sees and what the numbers cannot show.
A laboratory's Chemical Pathologist, Managers and IT can ask for the methods in full. That includes their parameters and EQAWarden's defaults. Calculated findings support the review; the interpretation and the recorded outcome are the reviewer's.
Result scoring against the analytical performance specification
RCPAQAP sets an analytical performance specification (APS) for each analyte. The APS score shows how much of that allowance a result has used, and on which side of the target.
The zones
- Withinwell inside the allowancewithin
- Warningwithin the APS but close to its limit. EQAWarden's zone, set at an ISO 13528 bandwarning
- Outsidebeyond the APS. RCPAQAP flags the result for reviewoutside
- Not scoredreported as less than or greater than, or no target or APSnot scored
RCPAQAP's assessment stands
RCPAQAP's own assessment is within or outside the APS. A result exactly at the limit keeps RCPAQAP's assessment.
- The Warning zone is EQAWarden's interpretation within the APS
- It is shown separately from RCPAQAP's categories
What one result can show. One result outside the APS warrants investigation, but does not by itself show whether the error is random or systematic.
A laboratory performing well will still see the occasional result in the Warning zone by chance.
Target reliability and uncertainty
RCPAQAP's expected result may be an all-method median, a category median or a specified target.
- Before the score is read, EQAWarden rates how far the target can be trusted
- The rating is adequate, limited or poor
- The rating describes the support for the target; it does not establish the true value
Small groups
A target from few laboratories is flagged and rated accordingly. The score is still calculated and still counts.
Target uncertainty
Where the report gives what is needed, the target's own uncertainty is estimated as ISO 13528 describes. A score is not interpreted where the target is too uncertain to support it.
Commutability
EQAWarden does not assume that RCPAQAP's material behaves like patient samples. All-method comparisons are therefore treated as descriptive. The conclusion that can be drawn is agreement with same-method peers.
Clerical error screening
Transcription is among the most frequent causes of EQA failure in the published troubleshooting reviews. It is also an early step in RCPAQAP's own interpretation flowchart.
- The screens run first, on results outside the APS only
- They never alter a result
- They look for a possible sample swap or a unit or decimal slip
- They also look for a digit error or a result entered in the wrong field
- Each candidate is to be checked against the record
- A candidate that fits better does not prove the error occurred
Series and cycle analysis
A single survey cannot separate bias from random error. A series is one analyte on one instrument over time. Across a series, EQAWarden looks at:
- Repeated results outside the APS
- Results sitting consistently above or below target
- Drift, runs on one side and sudden steps
Over the cycle it follows RCPAQAP's own regression of result on target, and checks:
- Whether the error is proportional or constant
- Whether imprecision meets the specification
Each check shows the minimum data it needs.
Many analytes, one report. A report may carry dozens of analytes, so some patterns appear by chance alone.
- EQAWarden allows for the number of analytes reviewed
- A pattern that does not survive that allowance is still shown, worded as possible
- RCPAQAP's single-result flags are never adjusted
Results before and after each recorded lot change
The lots put into use and their dates come from LotWarden or another lot source.
- EQAWarden works out which lots were in use on each analysis date
- It shows how confident that link is: high, medium or low
- Results before and after each reagent or calibrator lot change are compared
- Same-direction steps across analytes that share a reagent or calibrator are strong evidence of an association with that change
A step shows an association with the recorded lot change; it does not establish the lot as the cause. With few surveys either side, a comparison that finds nothing is never read as "no effect".
Comparison of IQC and EQA findings
For each control level and lot, the internal quality control (IQC) before the analysis date is compared with:
- The IQC on and just after it
- The EQA deviation
IQC that moved with the EQA result supports investigation of an analytical change, to be confirmed with patient samples.
IQC that did not move may point to a cause specific to the EQA samples, but does not by itself establish it.
Also reported in their own right:
- Calibrations and maintenance around the run
- Control lot changes and peer statistics around the run
- Rule violations recorded in Unity, where the export carries them
Evidence weighting and suggested outcomes
Potential causes are ranked by the evidence the findings give, to order what to check.
- The ranking is not a probability
- One pattern is never counted twice under different headings
- The suggested outcome comes from four fixed rules applied in order
- The reviewer records the outcome
Wording by strength of evidence
Finding text does not state a cause as fact. Instead, it:
- Gives the estimate and the count
- Names the target source
- Separates what was observed from its interpretation
- Lists the explanations that fit, with the check that tells them apart
Weak or too little
Finding text uses one of these:
- "is consistent with"
- "may indicate"
- "there is insufficient information to assess"
A clear signal
"suggests".
Corroborated
"provides strong evidence of", where independent records agree. Never "proves".
Small samples and their limits
A cycle gives only a few results for each analyte. Where the numbers cannot support a conclusion, the assessment says so.
One survey
Two results cannot separate bias from random error.
A lot change
With few surveys either side, "the lot change had no effect" cannot be concluded.
An SD, a line
An SD from few results is itself uncertain, and linearity cannot be judged from a handful of targets.
Sources of methods and thresholds
Every threshold the engine uses has a recorded source and reason.
- RCPAQAP, an international standard or a published method may set the number
- EQAWarden then uses it as published
- A published rule applied in a changed form is marked as adapted
- Where no source sets the number, it is an EQAWarden default: a design choice
- Each default is listed for the laboratory's Chemical Pathologist to ratify or change
- A senior scientist they delegate may do this instead
- That is done before EQAWarden's suggestions are used in review records
Every threshold, window and weight the engine reads. Each has its source and reason.
Design choices where no scheme or standard sets the number, listed in the sign-off pack.
Each is agreed or changed and then signed off. This happens before EQAWarden's suggestions go into the laboratory's review records.
Each published source is named beside the threshold it sets. The sources include:
- RCPAQAP and ISO 13528
- IUPAC and NIST
- Westgard, Bio-Rad and Fraser
How the laboratory ratifies the defaults
1. The sign-off pack
The sign-off pack lists each default with its reason and an Agree or Change column.
2. Ratified and signed
The laboratory's Chemical Pathologist agrees or changes each one. A senior scientist they delegate may do so instead.
- They sign the pack before suggestions go into review records
- The program records the status as Ratified, or as Ratified for an earlier release
- Otherwise the status is Not yet ratified
- It marks reviews with that status, and does not block a review
3. Changed under control
A change to an engine default is made in the knowledge files with a new version, never on a running copy.
- The versions are recorded on every assessment
- The laboratory's own weekly QC limits and settings are changed only by an Authoriser at Manager or Chemical Pathologist level
- Each such change needs a reason and an electronic signature
The defaults are tested starting values for the laboratory to ratify or change. They are not an external standard. The reviewer records the outcome of each review.
How they sit within the rest of the validation evidence: Validation.
Review the methods in a demonstration
The numbers on an assessment page link to the rule and the reference behind them. A demonstration can go through the methods with the laboratory's Chemical Pathologist, and the full description is given on request.