Standards alignment and supporting records
LotWarden records information relevant to reagent and consumable receipt, acceptance and traceability. Its assessment outputs organise available records against the requirements identified for the configured laboratory setting.
- ISO 15189:2022 for a pathology laboratory assessed by NATA
- ISO/IEC 17025:2017 for a forensic or testing laboratory
- ISO 15189 with the NPAAC requirements for a haemopoietic stem cell collection service
- The Australian Code of GMP for a cell therapy, cord blood or tissue laboratory
Under ISO 15189:2022:
- Clause 6.6 bears on what is recorded when reagents and consumables arrive
- Clause 7.6 bears on information systems and how the software itself is validated
- Clause 8.7 bears on how nonconformities and corrective actions are recorded
Assessment and audit packs are generated from these records without changing them. Out-of-service dating and recall actions remain the laboratory's own process.
The scope of each mapping, the records LotWarden supplies and the evidence the laboratory provides separately are described below.
Records made at receipt
Reception records
Each confirmed receipt is written to the working copy of the receiving department's log.
- The entry holds the date, lot and expiry
- It also holds the quantity, supplier and delivery note number
- The receiver is the signed-in person, recorded in Received By in Chemistry logs and in Comments elsewhere
- Staff transfer the entries into the controlled log as a separate step
Acceptance before use
Under the pathology setting, by default Chemistry lots are checked at receipt against the laboratory's pre-acceptance testing (PAT) register.
- An investigation worklist is generated
- A product the register lists needs PAT
- Any other product needs PAT unless it is ticked Exempt from PAT, with initials and a reason
- Other departments flag each new lot, confirm it before it is written and confirm bench verification
Condition on reception
The operator records the condition, including intact packaging and cold chain. It is written to the logs' Condition column. Failures are highlighted in red and recorded in the audit trail.
Expiry control
Expiry is checked on arrival and on the shelf.
- Stock expired on arrival is flagged and listed on the worklist with a quarantine action
- Short-dated stock is flagged amber
- Expired lots on the shelf are listed for removal
- Each department sets when a delivery counts as short-dated (90 days unless changed)
- It also sets when a lot on the shelf counts as expiring soon (30 days unless changed)
- A department can hold a supplier to the minimum expiry agreed with it
- Each change needs a reason, is made by an Authoriser and is audited
Supplier evaluation
For the laboratory's yearly supplier evaluation, LotWarden summarises each supplier's last 12 months:
- Deliveries
- Short-dated and expired arrivals
- Condition and QC failures
- Order lead times
Software validation
Information systems: developer release verification, with a validation dossier issued per release.
- Covers intended use, risk controls and change control
- A release check re-run for every issue, including a fictional-data regression suite
The laboratory's validation for its intended use is separate and starts from this dossier.
Nonconformities and corrective action
The corrective and preventive action (CAPA) register records each nonconformity.
- Opened from a lot held at receipt or by hand
- Worked through a seven-part tick sheet
- The decision on a proposed item, the cause and the closure are each signed electronically
- When EQAWarden is linked, its CAPA items arrive over the link or as a checked file
- EQAWarden signs the accept or dismiss step, the cause and the effectiveness check
- Closure is LotWarden's
- The assessment pack carries a CAPA register sheet
The licensed setting determines the clause each item cites
One build serves all four settings. The site's licence names the settings it may choose.
- An Authoriser chooses the setting in force on the Records tab, with a reason
- It is recorded in the audit trail and printed in every pack
| Pathology | Forensic and testing | HSC collection | Cell therapy, cord blood, tissue | |
|---|---|---|---|---|
| Body and visit | NATA assessment | NATA assessment | NATA assessment | TGA inspection |
| Standard | ISO 15189:2022 | ISO/IEC 17025:2017 with the Legal (forensic) appendix | ISO 15189:2022 with the NPAAC requirements for haemopoietic progenitor cells (edition for the site to confirm) | The Australian Code of GMP for human blood, tissues and cellular therapy products. Also PIC/S Annex 11 and FACT-JACIE 8th edition |
| The before-use check | PAT, against the PAT register | Before-use check, in the acceptance record | Lot release check, in the materials record | Lot release check, in the materials record |
| Which products | Those the register lists, and any marked critical | Every department on the acceptance record | Every material critical unless marked otherwise | Every product critical unless marked otherwise |
| An unreleased lot at its Date in Use | Warned, with a reason | Refused | Refused | Refused |
| A rejected or embargoed lot | Refused a Date in Use in every setting | |||
| The pack | NATA assessment pack, ISO 15189 clauses | NATA assessment pack, ISO/IEC 17025 clauses. Items to bring: critical reagents checked before use and reference materials identified. Also the lot recorded in case records and proficiency testing reviewed | NATA assessment pack, ISO 15189 clauses. Items to bring: quarantine and release, and collection traceability where linked. Also recalls and investigations, and certificates. Also in the pack: the four HSC collection logs | TGA inspection pack with Code, Annex 11 and FACT clauses. Items to bring: quarantine and approved suppliers. Also certificates by lot, the laboratory's validation and retention |
| Retention | NPAAC requirements | NATA's Legal appendix and the laboratory's policy | The NPAAC requirements and the laboratory's own schedule, to confirm | 30 years for traceability, to confirm against the Code's own clause |
| The clause map | The established ISO 15189:2022 clause map | Written for the laboratory's quality manager to confirm against the standard's own text before a pack goes out; items still to confirm are marked in the pack | ||
Each setting in full, with what the licence carries and the limits of each: the four laboratory settings.
Build identification and release verification
Three things are kept distinct:
- The developer's verification of each release
- The laboratory's validation of the software for its intended use
- The records the laboratory presents at assessment
Re-run before every dossier issue, with its results in the dossier.
Marks itself DRAFT if anything disagrees.
A section changed without a new version shows amber.
Release identified
Each build carries these, embedded at build time:
- A release number
- A version for each of seven program sections
- The commit
They are shown in the footer and written into every audit record.
- A section changed without a new version is marked UNRELEASED
- Uncommitted code is marked MODIFIED
- Both show amber
Release check
The release check covers:
- Reading on fictional and real delivery notes
- Column detection, writes into copies and duplicates
- Pre-acceptance and routing
- A re-check of the section versions and write confinement
- A re-check of the origin guard and the network guard
Validation dossier
Generated from the build record, the executable's SHA-256 and the release results together.
- It marks itself DRAFT if they disagree
- It also does so if code changed after the build or if a check failed
- Each build is hash-verified when it is copied to the supplied USB stick
Answer keys measure the delivery-note reader.
- Staff can correct a review to match the paper and save it as an answer key
- Before a new reader is released, LotWarden's developer scores it against every key the developer holds
- It is scored on four fields: the code and lot, then the expiry and quantity
- A laboratory's keys stay on its licensed PC. The programs upload nothing
- At the laboratory, the Reader accuracy page measures reading on the laboratory's own delivery notes
- It prints rates, counts and confidence intervals for the validation file
One page sets out EQAWarden's test cases and threshold definitions. It also covers laboratory sign-off and change control: EQAWarden validation.
Assessment packs for NATA (ISO 15189 or ISO/IEC 17025), TGA (the Code of GMP) or ISO 9001:2015
Choose the visit and the dates and select Build the pack. LotWarden generates two files from the selected records.
- A presentation of the key charts and tables
- A workbook of every record behind it
- A pack organises records against the mapped requirements. It does not demonstrate compliance with a whole standard
Contents of the pack
- A checklist of the mapped requirements, each item with its clause.
- Each item is marked with one of five statuses:
- In order
- Check
- Partly covered
- Bring separately
- For information
- Receipts, lots received and lots put into use for the period.
- Removals and shelf counts for the period.
- The CAPA register and audit records for the period.
- The acceptance record where it is in use.
- The laboratory setting printed in the details.
- Rows to check highlighted, with the reason in their first column.
Scope of the pack
- The pack changes nothing in the logs. It is generated from the records; the logs remain the record.
- Products ticked Exempt from PAT are not flagged for PAT; they are listed as exempt on the PAT scope sheet.
- A blank QC-at-first-use column is not flagged when that QC is recorded in another system.
- Items marked Bring separately are for the laboratory to supply from its own systems.
Common questions, with the records that relate to them
Is the software validated for its intended use?
Each release is issued with a validation dossier. That is the developer's release verification.
- The dossier covers intended use, risk controls and change control
- It also holds the release check results
- The release shown in the footer and the executable's SHA-256 match section 1 of the dossier
Validation for the laboratory's intended use is the laboratory's own, and starts from the dossier.
Who entered this receipt, and when?
Both the log row and the audit log record it.
- The log row carries the signed-in receiver and a delivery note stamp
- The audit log carries the signed-in person and the Windows username
- It also carries the timestamp and the build for the write
- The audit record is hash-chained to the records around it
How do you know a lot was accepted before it was used?
For lots checked against the PAT register, the acceptance status on the day of receipt is recorded. The investigation worklist names the action required.
- First use is recorded through Date in Use
- A lot the register shows REJECTED is refused there
- For a lot the register does not show as accepted, Date in Use warns and asks for a reason
- That reason is recorded in the audit trail with the signed-in person
- An expired lot is refused
- So is a lot whose QC is recorded as not acceptable
- So is a lot rejected or embargoed in LotWarden's acceptance record
Can the record be altered after the fact?
The log workbooks are the laboratory's own controlled documents and remain so.
- LotWarden's audit log is hash-chained: editing, inserting or deleting a record breaks the chain
- The page reports it on the next load, so tampering is detectable rather than prevented
- Where audit anchors on the shared drive are used, a later check can test the trail
- It shows whether the trail was cut, changed or replaced
- Sign-offs are kept in sealed (encrypted) records
Can a lot be used before it is released?
LotWarden records use. It does not physically prevent it.
- The forensic, HSC collection and cell therapy settings refuse a Date in Use until the lot has been released
- The same applies to a critical product whose flag is set to refuse
- Release is by a second person where the setting or the product's flag requires it
- The refusal is listed with its reason
- Otherwise the pathology setting warns for a lot not yet released or not accepted on the register
- A reason is needed
- A rejected or embargoed lot is refused a Date in Use in every setting
Is LotWarden TGA approved, GMP certified or ISO certified?
No.
- LotWarden is not registered or approved by the TGA and is not GMP or ISO certified
- NATA assesses laboratories, not software
- LotWarden supplies records relevant to the requirements mapped for the laboratory's setting
The laboratory validates it for its intended use, starting from the validation dossier issued with each release.
Where does the data go?
The data stays on the licensed PC. In the network version, it stays in the approved folders on the laboratory's shared drive.
- The programs serve their pages on the loopback address only
- A socket-level guard refuses their outbound connections and logs each attempt
- Sealed records are encrypted at rest
- Exports, generated sheets and audit logs are readable files that go only where a person saves them
Previous assessment findings and LotWarden's records
Bring your last assessment findings to a demonstration. We can show which LotWarden records relate to each finding, and what remains for the laboratory to supply.