EQAWarden · the rules

Documented assessment rules, in summary

EQAWarden applies 59 finding rules and 4 outcome rules to each RCPAQAP external quality assessment (EQA) survey. This page says what the rules look at, what the reviewer sees and what the reviewer decides.

  • 26 causes ranked by the evidence the findings give
  • The full rules catalogue on request: each rule's conditions, thresholds and sources
  • Given to a laboratory's Chemical Pathologist, Managers and IT

The principles behind every rule.

  • RCPAQAP's own assessment decides whether a result is within or outside the analytical performance specification (APS)
  • EQAWarden does not re-grade it, and RCPAQAP's terms are used on screen
  • Statistics describe patterns and do not establish a cause
  • Each finding shows its figures and target source
  • An assessment can be reproduced later from the same data and versions
What the rules look at

Seven groups of finding rules

Each finding names what was seen, shows the figures behind it and links to its references. The reviewer reads the findings: the rules do not decide the outcome.

Result and target

Each result is scored against its APS, high or low.

  • A result close to the limit but within it is shown in EQAWarden's Warning zone, apart from RCPAQAP's own categories
  • RCPAQAP's red or amber colour is shown
  • Results reported as less than or greater than are kept but not scored
  • The target itself is rated for how far it can be trusted
  • The rating is lower, for example when few laboratories reported or the analyte's methods are not standardised

Clerical screens

Tested only on results outside the APS. A transcription, unit or sample error would distort later statistics. So these screens run first. They look for:

  • A sample swap
  • A unit or decimal slip
  • A transposed or dropped digit
  • A result entered against another analyte

Survey and cross-analyte

Patterns across the analytes of one survey:

  • One vial shifted, as after a reconstitution error
  • Both vials shifted the same way
  • One instrument off
  • A set of analytes moving together as a known fault would move them

Series and cycle

One analyte on one instrument over time:

  • Repeated results outside the APS
  • Results sitting consistently above or below target
  • Drift, and runs on one side
  • Over the cycle, whether the error is proportional or constant
  • Whether imprecision meets the specification

Lots

Results before and after each reagent or calibrator lot change recorded in LotWarden or Unity.

  • How confidently each lot is linked to the results
  • A lot shared by several analytes
  • A lot put into use shortly before analysis, or used past its expiry
  • A control lot change

Internal QC

Whether the laboratory's internal quality control (IQC) moved with the EQA result or stayed steady. Also the events and peer statistics around the run:

  • Calibrations
  • Maintenance
  • Peer statistics
  • Rule violations, where Unity's export records them

Method group and knowledge

The laboratory's position within its method group, and what is known about the analyte.

  • Whether the method group itself reads away from the target
  • RCPAQAP's ranking of the laboratory
  • The known pitfalls in the analyte's monograph
  • A correlation factor set on the assay
Suggested outcome

Outcome rules and the suggested outcome

Four outcome rules give the suggested outcome, one of:

  • Corrective action
  • Investigate
  • Acceptable, monitor: within limits but with a pattern worth watching in later surveys
  • Acceptable

How they are applied:

  • Checked from the most serious down
  • The first that applies is shown with the findings behind it

The reviewer records the outcome. An outcome that differs from the suggestion asks for a reason.

The 26 causes

Potential causes ranked by supporting evidence

Each finding counts for or against the causes it bears on. The ranking orders what to check first.

  • A lot finding counts for less when the lot link is uncertain
  • Opening a cause shows the evidence for and against
  • The ranking is not a probability and does not establish a cause

Each cause is given one of four ratings:

  • Strongly supported
  • Supported
  • Possible
  • Not supported

Clerical

This group covers:

  • Transcription
  • A unit or conversion error
  • A sample or result swap

Material

This group covers:

  • Reconstitution or volume
  • Sample handling or stability
  • A matrix effect or method-group difference

Analytical

This group covers:

  • Calibration
  • A reagent or calibrator lot change
  • An instrument fault and imprecision

Also analyser faults, such as:

  • Reaction temperature, the ISE module and sample volume
  • Contamination and carry-over
  • An immunoassay signal fault
  • Software and sample pretreatment

Target

An uncertain or unsuitable target.

Chance

Chance variation, where nothing else fits.

Result path

This group covers:

  • A correlation factor applied to EQA results
  • A mapping, unit or rounding error in the middleware or LIS

For the likelier causes, the assessment lists the checks to make in order. Each check has its reason and references. The investigation playbooks.

  • Fuller technical detail of the rules, the outcome rules and the cause ranking is given on request
  • It goes to a laboratory's Chemical Pathologist, Managers and IT
  • The sign-off pack lists every EQAWarden default for ratification

Ask for the technical detail.

See the rules applied to a demonstration year

The demonstration uses a fictional year of survey data. It includes a reagent lot step and a sample swap. It also includes a unit error and a reconstitution error.