Getting ready for a NATA audit: a checklist for medical laboratories
NATA calls its visit an assessment. Most laboratory staff call it an audit. This checklist lists the records an assessment against ISO 15189:2022 looks at, and where each one comes from.
- Grouped by the clauses of ISO 15189:2022
- Each record shown with the system that usually keeps it
- LotWarden and EQAWarden keep the reagent records, the IQC and EQA review and the CAPA items that follow
- The rest come from the laboratory's other systems
- No single program gets a laboratory ready for a whole assessment
Assessors look at the records and watch the work
NATA accredits medical laboratories against ISO 15189 and the NPAAC requirements. An assessment checks that the laboratory's records and practice meet them.
- It covers the scope of accreditation: the examinations the laboratory is accredited for
- Technical assessors usually work in the same discipline as the laboratory
- They often trace one result back through its records
- That trace runs through the sample, the method, the reagent lot, the QC and the person
- Where a record or practice falls short, a finding is raised and the laboratory answers it with corrective action
A trace back from a result is the quickest test of readiness. If the laboratory can do it in minutes, most of the records are in order.
Records to have ready, by clause
The rows follow ISO 15189:2022. The last column names the system that usually keeps each record. Rows kept by LotWarden or EQAWarden say so.
| Area | Records to have ready | Usually kept in |
|---|---|---|
| Personnel (6.2) | Training and competence assessed for each person and examination. Authorisations to perform, review and release | The laboratory's training or quality system |
| Facilities and environment (6.3) | Room, refrigerator and freezer temperatures monitored. Action taken when a reading is out of range | The monitoring system or paper charts |
| Equipment and calibration (6.4 and 6.5) | Maintenance and service records. Calibration with its metrological traceability. Equipment taken out of service and returned | Equipment records and the analysers' own logs |
| Reagents and consumables (6.6) | Receipt with lot, expiry and condition. Acceptance before use. Date in Use. Expired stock removed | LotWarden: the department's logs, the lot check workbook and the assessment pack |
| Suppliers (6.8) | Approved suppliers and their yearly evaluation | The laboratory's own criteria. LotWarden supplies each supplier's last 12 months of deliveries as supporting records |
| Examination methods (7.3) | Verification or validation of each method. Measurement uncertainty where it applies | The laboratory's method files |
| Internal quality control (7.3.7.2) | IQC reviewed, with the action taken on each rejection recorded | The QC software. EQAWarden: the weekly QC review, signed by an Authoriser |
| External quality assessment (7.3.7.3) | Every survey reviewed and signed. Each unacceptable result investigated and acted on | EQAWarden: the survey record, cycle summary and NATA evidence pack |
| Information systems (7.6) | Validation of each program the laboratory relies on, for its intended use | The laboratory's validation files. LotWarden and EQAWarden each issue a validation dossier with every release |
| Documents and records (8.3 and 8.4) | Procedures current, reviewed and controlled. Records kept for their set periods | The document control system |
| Nonconformities and corrective action (8.7) | A register with each cause, action and effectiveness check | The QMS. LotWarden and EQAWarden: the CAPA register for lot, IQC and EQA items |
| Evaluations and internal audits (8.8) | Quality indicators. Internal audits covering the management system and the examinations | The quality system |
| Management review (8.9) | Minutes with the inputs reviewed and the actions agreed | The quality system |
| The last assessment | Each finding closed, with its evidence | The quality system |
Ten checks in the weeks before the visit
Most of the work is confirming that records exist and are complete. Work through these in order.
- Read the last assessment report and confirm each finding is closed, with its evidence
- Confirm every accredited examination has a current procedure, a verification and an EQA enrolment
- Pick a few recent results and trace each one back through its records, as an assessor would
- Check every reagent lot in use was received, accepted before use and given a Date in Use
- Take expired stock off the shelves and record the removal
- Confirm every EQA survey in the cycle is reviewed and signed, with its corrective actions closed
- Confirm IQC has been reviewed, with the action on each rejection recorded
- Bring the nonconformity register up to date, with effectiveness checks recorded
- Check internal audits and the management review were held and minuted within the cycle
- Check training and competence records are current for each person and examination
What LotWarden and EQAWarden prepare for the visit
Four rows of the checklist come mostly from the two programs' own records. Each output is built without changing any record.
LotWarden assessment pack
Choose NATA assessment (ISO 15189), set the period and build it. It gives:
- A presentation of the key charts and tables
- A workbook of every record, with rows to check highlighted
- A checklist by clause, each item marked In order, Check, Partly covered, Bring separately or For information
EQAWarden reports
For RCPAQAP surveys and internal QC:
- The survey record, with each result scored and the reviewer's signed outcome
- The cycle summary across the whole cycle
- The NATA evidence pack
- The weekly QC review, signed by an Authoriser
CAPA register
Items open from the records that found the problem:
- A lot held at receipt in LotWarden
- A rule breached in a signed week of IQC, or an EQA outcome signed as corrective action, in EQAWarden
- Each decision, cause and closure signed electronically
- A CAPA register sheet in LotWarden's assessment pack
More: Records the programs supply, CAPA register and Stock control software.
What no software does for you
Software can gather records. It cannot stand in for the laboratory at an assessment.
- Assessors watch the work and talk to staff, so records alone do not carry an assessment
- The laboratory sets its own acceptance criteria and decides each outcome
- NATA assesses laboratories, not software, and does not approve any program
- The laboratory validates each program it relies on for its intended use
- LotWarden and EQAWarden are not affiliated with or endorsed by NATA or ISO
Questions about getting ready for a NATA audit
How do I get ready for a NATA audit?
Start with the last assessment report and close each finding with its evidence. Then work through the records by clause of ISO 15189:2022. Three checks find most gaps:
- Trace a few recent results back through their records
- Check every reagent lot in use was accepted and given a Date in Use
- Confirm every EQA survey and IQC rejection was reviewed and acted on
Is a NATA assessment the same as an audit?
People use the two words for the same visit. NATA's own word is assessment.
An internal audit is different. It is the laboratory's own check of itself under clause 8.8 of ISO 15189:2022, and an assessor will ask to see it.
Is there software that gets a laboratory ready for a NATA audit?
No program covers a whole assessment. A quality management system covers documents, nonconformities, internal audits and training.
LotWarden and EQAWarden cover the reagent records, the IQC and EQA review and the CAPA items that follow from them. LotWarden builds an assessment pack with a checklist by clause, and marks what to bring from other systems.
What do NATA assessors look at for reagents?
They follow lots through clause 6.6 of ISO 15189:2022:
- The record of receipt, with lot, expiry and condition
- Acceptance before use
- The date each lot was put into use
- Expired stock removed from use
An assessor may pick a lot on the bench and ask for its records.
Does this checklist apply to an ISO/IEC 17025 laboratory?
The areas are much the same, but the clause numbers differ. This checklist follows ISO 15189:2022.
LotWarden's forensic setting maps its records to ISO/IEC 17025:2017 with the Legal (forensic) appendix.
Build the assessment pack on demonstration data
A 20-minute video call. We build the pack and open the EQA records on fictional data.