Triglyceride EQA results: interpretation and investigation guidance
The RCPAQAP limits for triglyceride, known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the triglyceride monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Check transcription and units (mmol/L versus mg/dL, ratio about 88.5).
Limits, units and performance goals for triglyceride
| RCPAQAP APS, General Serum Chemistry (GSC) | ±0.2 mmol/L up to 1.6 mmol/L, ±12% above |
|---|---|
| Units | mmol/L. 1 mg/dL = 0.011294 mmol/L. Factor as triolein (M 885.4 g/mol), C2 6.3. |
| Standardisation | Traceable to a reference method |
| Expected error mode | Proportional |
| Biological variation | Within-subject CV 19.76% and between-subject CV 35.3%, from the EFLM database (meta-analysis, 42 studies). |
| Desirable performance | Imprecision 9.88%, bias 10.11%, total allowable error 26.42%, derived from biological variation (Fraser). |
The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.
Known sources of error in triglyceride EQA results
In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.
Calibrator shared by several assays
Where one calibrator material serves several assays on an analyser, a reconstitution or volume error, an aged or wrongly stored vial, or a lot value entry error moves every assay calibrated from it by a similar percentage, at staggered dates when the assays are recalibrated at different times.
Free glycerol in the material read as triglyceride
Glycerol-based triglyceride methods measure glycerol released from triglyceride plus any free glycerol; without a free-glycerol blank, glycerol present in processed material (for example as a stabiliser) is reported as triglyceride.
Peroxidase (Trinder) end-point interference from reductants
Triglyceride is measured with a peroxidase-coupled colour end point; reductants such as ascorbate, N-acetylcysteine and metamizole consume peroxide or the dye before it is read and depress the result.[1, 2]
Non-commutable processed material
In a commutability study of 32 processed materials measured by 14 routine triglyceride methods, EQA/PT materials, calibrators and controls showed method-dependent matrix effects, while frozen serum and swine sera were commutable; aqueous glycerol solutions were generally non-commutable[3].
Lipoprotein stratification after freeze-thaw or incomplete mixing
Triglyceride-rich lipoproteins float and denature; an unmixed or partly thawed vial gives unrepresentative aliquots.
Reconstitution volume error or evaporation of the EQA vial
Too much or too little diluent at reconstitution, or water loss from an open or poorly stoppered vial, scales every dilution-linear analyte in that vial by the same percentage.
Correlation factor, unit conversion or LIS mapping in the result path
A correlation slope or intercept saved in the analyser's test settings, or a conversion in the middleware or LIS, is applied to every result for that assay, EQA included; RCPAQAP expects EQA results without such factors.[4]
Investigation checks in suggested order
- Check transcription and units (mmol/L versus mg/dL, ratio about 88.5).
- Check H and I indices, vial mixing and whether the result came from an auto-repeat dilution.
- Compare with the same-method peer group for triglyceride.
- Look for concordant shifts in the other assays calibrated from the same calibrator and the other peroxidase end-point analytes of the same sample.
- Check for a correlation factor or unit conversion in the analyser's test settings, the middleware or the LIS.
Unit traps
- Ratio about 88.5 (or 0.0113) = mg/dL versus mmol/L confusion (factor as triolein).
EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs
The full triglyceride monograph in EQAWarden
This page is a summary. The monograph EQAWarden ships with holds 24 entries on what can go wrong with triglyceride in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each triglyceride result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about triglyceride opens it.
Questions about triglyceride EQA
What is the RCPAQAP analytical performance specification for triglyceride?
In the General Serum Chemistry (GSC) programme it is ±0.2 mmol/L up to 1.6 mmol/L, ±12% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.
What is a common cause of a triglyceride EQA result outside the APS?
Calibrator shared by several assays. Where one calibrator material serves several assays on an analyser, a reconstitution or volume error, an aged or wrongly stored vial, or a lot value entry error moves every assay calibrated from it by a similar percentage, at staggered dates when the assays are recalibrated at different times.
What should be checked first when a triglyceride EQA result is outside the APS?
Check transcription and units (mmol/L versus mg/dL, ratio about 88.5).
How are triglyceride units converted?
1 mg/dL = 0.011294 mmol/L.
Where every fact on this page came from.
- Martinello F, Luiz da Silva E. Mechanism of ascorbic acid interference in biochemical tests that use peroxide and peroxidase to generate chromophore. Clin Chim Acta 2006;373:108-116.
- Spain MA, Wu AH. Bilirubin interference with determination of uric acid, cholesterol, and triglycerides in commercial peroxidase-coupled assays, and the effect of ferrocyanide. Clin Chem 1986;32:518-521.
- Meng Q, Zhou W, Zhang C, et al. Serum triglyceride measurements: the commutability of reference materials and the accuracy of results. Clin Chem Lab Med 2017;55.
- myQAP help. Post Analytical Correction Factors (Wayback 2026-01-13).
- RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/