EQAWarden knowledge base · NT-proBNP

NT-proBNP EQA results: interpretation and investigation guidance

The RCPAQAP limits for NT-proBNP, known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the NT-proBNP monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: For a high sample, check the dilution and its recovery; for a low sample, check for a result entered as below the reporting limit.

At a glance

Limits, units and performance goals for NT-proBNP

RCPAQAP APS, B-Natriuretic Peptide±25 ng/L up to 125 ng/L, ±20% above
Unitsng/L. 1 pg/mL = 1 ng/L
StandardisationNot standardised: compare with the method group
Expected error modeProportional

The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.

Known EQA pitfalls

Known sources of error in NT-proBNP EQA results

In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.

Method-group target: NT-proBNP assays are not standardised

Assays use different antibodies and internal standards, cross-react differently with proBNP and miss glycosylated forms, so method groups separate on the same material.[1, 2]

proBNP and glycosylated forms in the EQA material

NT-proBNP assays differ in cross-reaction with proBNP and do not detect glycosylated NT-proBNP, so the response to an EQA material depends on which peptide forms it contains, which is not published.[1, 3]

New reagent lot (NT-proBNP)

Each reagent lot can carry its own lot-specific calibration curve; a lot-to-lot difference that recalibration does not remove moves every result on every analyser using that lot.[4]

Calibrator handling at recalibration (NT-proBNP)

A reconstitution volume error, an evaporated or out-of-date reconstituted calibrator, or a calibrator from the wrong lot shifts every result after that calibration on that analyser.

Chemiluminescent signal or wash fault in a sandwich assay

In a two-site sandwich assay the signal rises with the analyte.[5]

High-dose hook effect (sandwich format)

Two-site sandwich assays can under-read at extreme concentrations when the antibodies are saturated.[6, 7]

Heterophile or human anti-animal antibodies

Immunoassays are designed to minimise heterophile antibody interference, but such antibodies can still give falsely high or low results.[8, 9]

Investigation checks

Investigation checks in suggested order

  1. For a high sample, check the dilution and its recovery; for a low sample, check for a result entered as below the reporting limit.
  2. If BNP from the same material is also low, consider BNP-specific degradation or material handling.
  3. Check units: pg/mL = ng/L; pmol/L = pg/mL x 0.118.
  4. Confirm which NT-proBNP assay each analyser runs and that the RCPAQAP method code matches; compare with that method group.
  5. Check calibration and reagent lot dates and calibrator age.
  6. If several sandwich immunoassays on one analyser moved together, review signal reagent, wash and detector records.

Unit traps

  • pg/mL and ng/L are numerically identical.

EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs

In EQAWarden

The full NT-proBNP monograph in EQAWarden

This page is a summary. The monograph EQAWarden ships with holds 34 entries on what can go wrong with NT-proBNP in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each NT-proBNP result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about NT-proBNP opens it.

Plainly answered

Questions about NT-proBNP EQA

What is the RCPAQAP analytical performance specification for NT-proBNP?

In the B-Natriuretic Peptide programme it is ±25 ng/L up to 125 ng/L, ±20% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.

What is a common cause of an NT-proBNP EQA result outside the APS?

Method-group target: NT-proBNP assays are not standardised. Assays use different antibodies and internal standards, cross-react differently with proBNP and miss glycosylated forms, so method groups separate on the same material.

What should be checked first when an NT-proBNP EQA result is outside the APS?

For a high sample, check the dilution and its recovery; for a low sample, check for a result entered as below the reporting limit.

How are NT-proBNP units converted?

1 pg/mL = 1 ng/L.

References

Where every fact on this page came from.

  1. Saenger AK, Rodriguez-Fraga O, Ler R, Ordonez-Llanos J, Jaffe AS, Goetze JP, Apple FS. Specificity of B-type natriuretic peptide assays: cross-reactivity with different BNP, NT-proBNP, and proBNP peptides. Clin Chem 2017;63:351-358.
  2. Badrick T, Miller WG, Panteghini M, Delatour V, Berghall H, MacKenzie F, Jones G. Interpreting EQA - understanding why commutability of materials matters. Clin Chem 2022;68:494-500. doi:10.1093/clinchem/hvac002
  3. Luckenbill KN, Christenson RH, Jaffe AS, et al. Cross-reactivity of BNP, NT-proBNP, and proBNP in commercial BNP and NT-proBNP assays: preliminary observations from the IFCC Committee for Standardization of Markers of Cardiac Damage. Clin Chem 2008;54:619-621 (letter; abstract not available, not read in full).
  4. Miller WG, Erek A, Cunningham TD, Oladipo O, Scott MG, Johnson RE. Commutability limitations influence quality control results with different reagent lots. Clin Chem 2011;57:76-83. doi:10.1373/clinchem.2010.148106
  5. Wild D (ed.). The Immunoassay Handbook: Theory and Applications of Enzyme Immunoassay, Electrochemiluminescence and Related Techniques. 4th ed. Oxford: Elsevier; 2013.
  6. Jassam N, Jones CM, Briscoe T, Horner JH. The hook effect: a need for constant vigilance. Ann Clin Biochem 2006;43:314-317. doi:10.1258/000456306777695726
  7. Dodig S. Interferences in quantitative immunochemical methods. Biochem Med (Zagreb) 2009;19:50-62. doi:10.11613/BM.2009.005
  8. Kricka LJ. Human anti-animal antibody interferences in immunological assays. Clin Chem 1999;45:942-956. doi:10.1093/clinchem/45.7.942
  9. Tate J, Ward G. Interferences in immunoassay. Clin Biochem Rev 2004;25:105-120 (AACB journal). https://pmc.ncbi.nlm.nih.gov/articles/PMC1904417/
  10. RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/