LDL cholesterol EQA results: interpretation and investigation guidance
The RCPAQAP limits for LDL cholesterol, known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the LDL cholesterol monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Establish whether the submitted LDL is direct or calculated, and which equation; check the registered method.
Limits, units and performance goals for LDL cholesterol
| RCPAQAP APS, Special Lipids | ±0.2 mmol/L up to 2 mmol/L, ±10% above |
|---|---|
| Units | mmol/L. 1 mg/dL = 0.025862 mmol/L |
| Standardisation | Partly standardised |
| Expected error mode | Mixed |
| Biological variation | Within-subject CV 7.43% and between-subject CV 24.33%, from the EFLM database (meta-analysis, 21 studies). |
| Desirable performance | Imprecision 3.71%, bias 6.36%, total allowable error 12.49%, derived from biological variation (Fraser). |
The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.
Known sources of error in LDL cholesterol EQA results
In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.
Direct LDL on lyophilised or frozen processed material
Non-linearity of LDL cholesterol on lyophilised material has been attributed to a matrix effect, and none of four CDC frozen pools was commutable for direct LDL methods under medical-requirement criteria; results on processed material therefore separate by method.[1, 2, 3]
Calculated versus direct LDL in the same comparison
A calculated LDL inherits the errors of total cholesterol, HDL-C and triglyceride and depends on the equation (Friedewald, NIH/Sampson); it differs from direct methods especially at high triglyceride.[4, 5]
Formula, unit or constant error in a calculated LDL
A calculated test built on the analyser, in middleware or in the LIS can carry its own formula, divisor (5 for mg/dL versus 2.2 for mmol/L) or unit error.[4]
Peroxidase (Trinder) end-point interference from reductants
Direct LDL methods with a peroxidase-coupled colour end point can be depressed by reducing substances such as N-acetylcysteine that consume hydrogen peroxide or the oxidised chromogen.[6, 7]
Reconstitution volume error or evaporation of the EQA vial
Too much or too little diluent at reconstitution, or water loss from an open or poorly stoppered vial, scales every dilution-linear analyte in that vial by the same percentage.
Correlation factor, unit conversion or LIS mapping in the result path
A correlation slope or intercept, unit conversion or result mapping saved in the analyser, middleware or LIS is applied to every result for that test, EQA included; RCPAQAP expects EQA results without such factors.[8]
Investigation checks in suggested order
- Establish whether the submitted LDL is direct or calculated, and which equation; check the registered method.
- If calculated, recompute from the submitted TC, HDL-C and TG and examine each input's deviation first.
- Check H, I and L indices of the sample.
- Check for a correlation factor, ratio-test formula or middleware calculation change.
- If direct, compare with the peer group of the direct LDL method in use; review calibrator reconstitution, calibration dates and reagent lot changes.
Unit traps
- Ratio about 38.7 = mg/dL versus mmol/L confusion.
- Friedewald divisor: TG/5 applies in mg/dL and TG/2.2 in mmol/L; a formula using the wrong divisor gives a triglyceride-dependent error.
EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs
The full LDL cholesterol monograph in EQAWarden
This page is a summary. The monograph EQAWarden ships with holds 26 entries on what can go wrong with LDL cholesterol in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each LDL cholesterol result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about LDL cholesterol opens it.
Questions about LDL cholesterol EQA
What is the RCPAQAP analytical performance specification for LDL cholesterol?
In the Special Lipids programme it is ±0.2 mmol/L up to 2 mmol/L, ±10% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.
What is a common cause of a LDL cholesterol EQA result outside the APS?
Direct LDL on lyophilised or frozen processed material. Non-linearity of LDL cholesterol on lyophilised material has been attributed to a matrix effect, and none of four CDC frozen pools was commutable for direct LDL methods under medical-requirement criteria; results on processed material therefore separate by method.
What should be checked first when a LDL cholesterol EQA result is outside the APS?
Establish whether the submitted LDL is direct or calculated, and which equation; check the registered method.
How are LDL cholesterol units converted?
1 mg/dL = 0.025862 mmol/L.
Where every fact on this page came from.
- Kroll MH, Chesler R. The nonlinearity seen for LDL-cholesterol with lyophilized material is a matrix effect. Clin Chem 1998;44:1770-1771.
- Korzun WJ, Nilsson G, Bachmann LM, et al. Difference in bias approach for commutability assessment: application to frozen pools of human serum measured by 8 direct methods for HDL and LDL cholesterol. Clin Chem 2015;61:1107-1113.
- Miller WG, Myers GL, Sakurabayashi I, et al. Seven direct methods for measuring HDL and LDL cholesterol compared with ultracentrifugation reference measurement procedures. Clin Chem 2010;56:977-986.
- Friedewald WT, Levy RI, Fredrickson DS. Estimation of the concentration of low-density lipoprotein cholesterol in plasma, without use of the preparative ultracentrifuge. Clin Chem 1972;18:499-502.
- Sampson M, Ling C, Sun Q, et al. A new equation for calculation of low-density lipoprotein cholesterol in patients with normolipidemia and/or hypertriglyceridemia. JAMA Cardiol 2020;5:540-548.
- Martinello F, Luiz da Silva E. Mechanism of ascorbic acid interference in biochemical tests that use peroxide and peroxidase to generate chromophore. Clin Chim Acta 2006;373:108-116.
- Spain MA, Wu AH. Bilirubin interference with determination of uric acid, cholesterol, and triglycerides in commercial peroxidase-coupled assays, and the effect of ferrocyanide. Clin Chem 1986;32:518-521.
- myQAP help. Post Analytical Correction Factors (Wayback 2026-01-13).
- RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/
See how EQAWarden reviews a LDL cholesterol result
Twenty minutes on a video call, on a fictional demo year: the score, the findings, the ranked causes and the signed review.