Lactate EQA results: interpretation and investigation guidance
The RCPAQAP limits for lactate, known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the lactate monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Check units (mmol/L versus mg/dL, ratio about 9.0) and transcription.
Limits, units and performance goals for lactate
| RCPAQAP APS, General Serum Chemistry (GSC) | ±0.5 mmol/L up to 4 mmol/L, ±12% above |
|---|---|
| RCPAQAP APS, Blood Gases | ±0.5 mmol/L up to 4 mmol/L, ±12% above |
| Units | mmol/L. 1 mg/dL = 0.11101 mmol/L |
| Standardisation | Not standardised: compare with the method group |
| Expected error mode | Proportional |
| Biological variation | Within-subject CV 30.5% and between-subject CV 19.3%, from the EFLM database (single-study, 1 study). |
| Desirable performance | Imprecision 15.25%, bias 9.02%, total allowable error 34.19%, derived from biological variation (Fraser). |
The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.
Known sources of error in lactate EQA results
In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.
Shared multi-analyte calibrator vial or lot
Where lactate is calibrated from a calibrator that also calibrates other assays, a reconstitution volume error, an aged or poorly stored vial or a value entry error moves every assay recalibrated from it by a similar percentage, at staggered dates when the calibration intervals differ.
Method-principle differences on processed material
Lactate methods differ in principle (photometric lactate oxidase or lactate dehydrogenase methods on chemistry analysers, amperometric sensors on blood gas analysers) and can differ systematically on processed material.
Peroxidase (Trinder) end-point interference from reductants
Lactate oxidase methods with a peroxidase-coupled colour end point can be depressed by reducing substances such as N-acetylcysteine; some reagents include ascorbate oxidase to remove ascorbate.[1, 2]
Change in lactate after reconstitution
Reconstituted lyophilised serum has no cells, so glycolysis is not expected, but microbial growth in a vial left open or warm could consume or produce lactate; the size in EQA material is not published.
Reconstitution volume error or evaporation of the EQA vial
Too much or too little diluent at reconstitution, or water loss from an open or poorly stoppered vial, scales every dilution-linear analyte in that vial by the same percentage.
Correlation factor, unit conversion or LIS mapping in the result path
A correlation slope or intercept, unit conversion or result mapping saved in the analyser, middleware or LIS is applied to every result for that test, EQA included; RCPAQAP expects EQA results without such factors.[3]
Investigation checks in suggested order
- Check units (mmol/L versus mg/dL, ratio about 9.0) and transcription.
- Check the time from reconstitution to analysis.
- Check H and I indices.
- Check the configured calibration interval and calibration history; check the calibrator age and whether other assays calibrated from it moved.
Unit traps
- Ratio about 9.0 (or 0.111) = mg/dL versus mmol/L confusion.
EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs
The full lactate monograph in EQAWarden
This page is a summary. The monograph EQAWarden ships with holds 25 entries on what can go wrong with lactate in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each lactate result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about lactate opens it.
Questions about lactate EQA
What is the RCPAQAP analytical performance specification for lactate?
In the General Serum Chemistry (GSC) programme it is ±0.5 mmol/L up to 4 mmol/L, ±12% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.
What is a common cause of a lactate EQA result outside the APS?
Shared multi-analyte calibrator vial or lot. Where lactate is calibrated from a calibrator that also calibrates other assays, a reconstitution volume error, an aged or poorly stored vial or a value entry error moves every assay recalibrated from it by a similar percentage, at staggered dates when the calibration intervals differ.
What should be checked first when a lactate EQA result is outside the APS?
Check units (mmol/L versus mg/dL, ratio about 9.0) and transcription.
How are lactate units converted?
1 mg/dL = 0.11101 mmol/L.
Where every fact on this page came from.
- Martinello F, Luiz da Silva E. Mechanism of ascorbic acid interference in biochemical tests that use peroxide and peroxidase to generate chromophore. Clin Chim Acta 2006;373:108-116.
- Spain MA, Wu AH. Bilirubin interference with determination of uric acid, cholesterol, and triglycerides in commercial peroxidase-coupled assays, and the effect of ferrocyanide. Clin Chem 1986;32:518-521.
- myQAP help. Post Analytical Correction Factors (Wayback 2026-01-13).
- RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/
See how EQAWarden reviews a lactate result
Twenty minutes on a video call, on a fictional demo year: the score, the findings, the ranked causes and the signed review.