EQAWarden knowledge base · Iron

Iron EQA results: interpretation and investigation guidance

The RCPAQAP limits for iron, known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the iron monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Check units (µmol/L versus µg/dL, ratio about 5.6) and transcription.

At a glance

Limits, units and performance goals for iron

RCPAQAP APS, General Serum Chemistry (GSC)±3 µmol/L up to 25 µmol/L, ±12% above
Unitsµmol/L. 1 µg/dL = 0.17905 µmol/L
StandardisationPartly standardised
Expected error modeProportional
Biological variationWithin-subject CV 25.24% and between-subject CV 27.99%, from the EFLM database (meta-analysis, 5 studies).
Desirable performanceImprecision 12.62%, bias 9.42%, total allowable error 30.25%, derived from biological variation (Fraser).

The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.

Known EQA pitfalls

Known sources of error in iron EQA results

In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.

Haemolysis raises iron

Red cells contain iron, so haemolysis raises measured serum iron, and free haemoglobin can also interfere spectrally; haemolysed samples are unsuitable for iron measurement.

Shared multi-analyte calibrator vial or lot

Where iron is calibrated from a calibrator that also calibrates other assays, a reconstitution volume error, an aged or poorly stored vial or a value entry error moves every assay recalibrated from it by a similar percentage, at staggered dates when the calibration intervals differ.

Iron contamination of the vial, diluent water or pipette tips

Iron is ubiquitous in glassware, water and some plastics; contamination adds a fixed amount of iron to the sample.

Liquid Serum Chemistry material from haemochromatosis donors

RCPAQAP LSC material is described as coming from venesection (haemochromatosis) donors, so iron-related concentrations may be atypical.[1]

Reconstitution volume error or evaporation of the EQA vial

Too much or too little diluent at reconstitution, or water loss from an open or poorly stoppered vial, scales every dilution-linear analyte in that vial by the same percentage.

Correlation factor, unit conversion or LIS mapping in the result path

A correlation slope or intercept, unit conversion or result mapping saved in the analyser, middleware or LIS is applied to every result for that test, EQA included; RCPAQAP expects EQA results without such factors.[2]

Investigation checks

Investigation checks in suggested order

  1. Check units (µmol/L versus µg/dL, ratio about 5.6) and transcription.
  2. Check the H and L indices and appearance of the sample.
  3. Review the calibration (calibrator lot, reconstitution date and calibration date), reagent pack age and IQC.
  4. Compare other analytes calibrated from the same calibrator and, for Liquid Serum Chemistry samples, transferrin, TIBC and ferritin of the same sample.

Unit traps

  • Ratio about 5.6 (or 0.179) = µg/dL versus µmol/L confusion.

EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs

In EQAWarden

The full iron monograph in EQAWarden

This page is a summary. The monograph EQAWarden ships with holds 24 entries on what can go wrong with iron in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each iron result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about iron opens it.

Plainly answered

Questions about iron EQA

What is the RCPAQAP analytical performance specification for iron?

In the General Serum Chemistry (GSC) programme it is ±3 µmol/L up to 25 µmol/L, ±12% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.

What is a common cause of an iron EQA result outside the APS?

Haemolysis raises iron. Red cells contain iron, so haemolysis raises measured serum iron, and free haemoglobin can also interfere spectrally; haemolysed samples are unsuitable for iron measurement.

What should be checked first when an iron EQA result is outside the APS?

Check units (µmol/L versus µg/dL, ratio about 5.6) and transcription.

How are iron units converted?

1 µg/dL = 0.17905 µmol/L.

References

Where every fact on this page came from.

  1. RCPAQAP product pages for General Serum Chemistry, Liquid Serum Chemistry, Alcohol and Ammonia, Blood Gases, Trace Elements: Serum (rcpaqap.com.au/product/...). Direct fetch returned HTTP 403 in this session; content (sample type, volume, storage on receipt, LSC 'commutable, haemochromatosis venesection source') was taken from search-engine summaries of the pages. Re-read the live pages.