HDL cholesterol EQA results: interpretation and investigation guidance
The RCPAQAP limits for HDL cholesterol, known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the HDL cholesterol monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Check the bilirubin target and I index of the sample, and the concentration (low HDL-C scatters more).
Limits, units and performance goals for HDL cholesterol
| RCPAQAP APS, General Serum Chemistry (GSC) | ±0.1 mmol/L up to 0.8 mmol/L, ±12% above |
|---|---|
| Units | mmol/L. 1 mg/dL = 0.025862 mmol/L |
| Standardisation | Partly standardised |
| Expected error mode | Mixed |
| Biological variation | Within-subject CV 5.74% and between-subject CV 22.8%, from the EFLM database (meta-analysis, 36 studies). |
| Desirable performance | Imprecision 2.87%, bias 5.88%, total allowable error 10.61%, derived from biological variation (Fraser). |
The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.
Known sources of error in HDL cholesterol EQA results
In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.
Direct HDL method on processed or lyophilised material
Direct homogeneous HDL assays depend on detergent and accelerator selectivity for lipoprotein particles; processing alters particles, so results on lyophilised material can differ between direct methods in ways not seen in patient samples.[1, 2, 3, 4]
Peroxidase (Trinder) end-point interference from reductants
Direct HDL cholesterol is measured with a peroxidase-coupled colour end point; reductants such as ascorbate, N-acetylcysteine and metamizole consume peroxide or the dye before it is read and depress the result.[5, 6]
HDL-C rise on standing at room temperature
An increase of HDL-C in serum stored at room temperature has been reported; the size in reconstituted EQA material is not known.[7]
Reconstitution volume error or evaporation of the EQA vial
Too much or too little diluent at reconstitution, or water loss from an open or poorly stoppered vial, scales every dilution-linear analyte in that vial by the same percentage.
Correlation factor, unit conversion or LIS mapping in the result path
A correlation slope or intercept saved in the analyser's test settings, or a conversion in the middleware or LIS, is applied to every result for that assay, EQA included; RCPAQAP expects EQA results without such factors.[8]
Investigation checks in suggested order
- Check the bilirubin target and I index of the sample, and the concentration (low HDL-C scatters more).
- If the laboratory calculates LDL, remember an HDL error passes into calculated LDL with the opposite sign.
- Confirm which HDL method was in use on the analysis date and that the registered method code matches it; compare with that peer group.
- Check whether total cholesterol and triglyceride of the same sample also deviate (shared vial or shared calibrator issue) or HDL alone (HDL calibrator, reagent or matrix).
Unit traps
- Ratio about 38.7 = mg/dL versus mmol/L confusion (same factor as cholesterol).
EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs
The full HDL cholesterol monograph in EQAWarden
This page is a summary. The monograph EQAWarden ships with holds 22 entries on what can go wrong with HDL cholesterol in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each HDL cholesterol result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about HDL cholesterol opens it.
Questions about HDL cholesterol EQA
What is the RCPAQAP analytical performance specification for HDL cholesterol?
In the General Serum Chemistry (GSC) programme it is ±0.1 mmol/L up to 0.8 mmol/L, ±12% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.
What is a common cause of a HDL cholesterol EQA result outside the APS?
Direct HDL method on processed or lyophilised material. Direct homogeneous HDL assays depend on detergent and accelerator selectivity for lipoprotein particles; processing alters particles, so results on lyophilised material can differ between direct methods in ways not seen in patient samples.
What should be checked first when a HDL cholesterol EQA result is outside the APS?
Check the bilirubin target and I index of the sample, and the concentration (low HDL-C scatters more).
How are HDL cholesterol units converted?
1 mg/dL = 0.025862 mmol/L.
Where every fact on this page came from.
- Miller WG, Myers GL, Sakurabayashi I, et al. Seven direct methods for measuring HDL and LDL cholesterol compared with ultracentrifugation reference measurement procedures. Clin Chem 2010;56:977-986.
- Korzun WJ, Nilsson G, Bachmann LM, et al. Difference in bias approach for commutability assessment: application to frozen pools of human serum measured by 8 direct methods for HDL and LDL cholesterol. Clin Chem 2015;61:1107-1113.
- EPTIS (BAM) database entry 183532: RCPAQAP Special Lipids (lyophilised serum).
- RCPAQAP Chemical Pathology. QAP Commutability & Matrix Effects. 2013.
- Martinello F, Luiz da Silva E. Mechanism of ascorbic acid interference in biochemical tests that use peroxide and peroxidase to generate chromophore. Clin Chim Acta 2006;373:108-116.
- Spain MA, Wu AH. Bilirubin interference with determination of uric acid, cholesterol, and triglycerides in commercial peroxidase-coupled assays, and the effect of ferrocyanide. Clin Chem 1986;32:518-521.
- Donnelly JG, Soldin SJ, Nealon DA, Hicks JM. Stability of twenty-five analytes in human serum at 22 C, 4 C, and -20 C. Pediatr Pathol Lab Med 1995;15:869-874.
- myQAP help. Post Analytical Correction Factors (Wayback 2026-01-13).
- RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/
See how EQAWarden reviews a HDL cholesterol result
Twenty minutes on a video call, on a fictional demo year: the score, the findings, the ranked causes and the signed review.