EQAWarden knowledge base · Gamma-glutamyltransferase (GGT)

Gamma-glutamyltransferase (GGT) EQA results: interpretation and investigation guidance

The RCPAQAP limits for gamma-glutamyltransferase (GGT), known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the gamma-glutamyltransferase (GGT) monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Line up reagent pack changes and calibrations with the EQA date.

At a glance

Limits, units and performance goals for gamma-glutamyltransferase (GGT)

RCPAQAP APS, General Serum Chemistry (GSC)±5 U/L up to 40 U/L, ±12% above
UnitsU/L. 1 ukat/L = 60 U/L
StandardisationTraceable to a reference method
Expected error modeProportional
Biological variationWithin-subject CV 8.29% and between-subject CV 45.1%, from the EFLM database (meta-analysis, 6 studies).
Desirable performanceImprecision 4.14%, bias 11.46%, total allowable error 18.3%, derived from biological variation (Fraser).

The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.

Known EQA pitfalls

Known sources of error in gamma-glutamyltransferase (GGT) EQA results

In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.

Method groups for GGT

Substrate, buffer and calibration differences leave some GGT methods 18-20% apart in commutable samples[1]; processed material can separate groups further[2].

Calibrator shared by several assays

Where one calibrator material serves several assays on an analyser, a reconstitution or volume error, an aged or wrongly stored vial, or a lot value entry error moves every assay calibrated from it by a similar percentage, at staggered dates when the assays are recalibrated at different times.

Reaction temperature away from 37 C

Catalytic activity rises with temperature, so every enzyme read in a cuvette that is warmer or cooler than 37 C is proportionally high or low.[3, 4]

Non-human or recombinant enzyme in processed EQA material

Lyophilised EQA material may carry enzyme of animal, recombinant or purified human origin whose response to substrate, activators and buffer differs from native human serum enzyme, so the material is not commutable for GGT and method groups separate in a way patient samples would not.[2, 5]

EQA vial reconstitution volume or evaporation

Under- or over-dilution when the lyophilised vial is reconstituted, or water loss from an open vial or sample cup, changes every solute and enzyme activity in that vial by the same proportion.[6]

Investigation checks

Investigation checks in suggested order

  1. Line up reagent pack changes and calibrations with the EQA date.
  2. Check the sample description for haemolysis.
  3. Check units (ratio 60 for ukat/L).
  4. Compare the other assays calibrated from the same calibrator and all enzymes (temperature) in the same survey.

Unit traps

  • Ratio 60 (or 0.0167) = ukat/L vs U/L.

EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs

In EQAWarden

The full gamma-glutamyltransferase (GGT) monograph in EQAWarden

This page is a summary. The monograph EQAWarden ships with holds 24 entries on what can go wrong with gamma-glutamyltransferase (GGT) in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each gamma-glutamyltransferase (GGT) result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about gamma-glutamyltransferase (GGT) opens it.

Plainly answered

Questions about gamma-glutamyltransferase (GGT) EQA

What is the RCPAQAP analytical performance specification for gamma-glutamyltransferase (GGT)?

In the General Serum Chemistry (GSC) programme it is ±5 U/L up to 40 U/L, ±12% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.

What is a common cause of a gamma-glutamyltransferase (GGT) EQA result outside the APS?

Method groups for GGT. Substrate, buffer and calibration differences leave some GGT methods 18-20% apart in commutable samples; processed material can separate groups further.

What should be checked first when a gamma-glutamyltransferase (GGT) EQA result is outside the APS?

Line up reagent pack changes and calibrations with the EQA date.

How are gamma-glutamyltransferase (GGT) units converted?

1 ukat/L = 60 U/L.

References

Where every fact on this page came from.

  1. Koerbin G, Tate JR, Ryan J, Jones GRD, Sikaris KA, et al. Bias assessment of general chemistry analytes using commutable samples. Clin Biochem Rev 2014;35:203-211.
  2. Miller WG, Jones GRD, Horowitz GL, Weykamp C. Proficiency testing/external quality assessment: current challenges and future directions. Clin Chem 2011;57:1670-80.
  3. Szasz G. The effect of temperature on enzyme activity and on the affinity of enzymes to their substrates. Z Klin Chem Klin Biochem 1974;12:166-170.
  4. Derived arithmetic (not a publication): per-degree temperature coefficients near 37 C computed as (F)^(1/7)-1 from published 37/30 C conversion factors, or from activation energies with the Arrhenius equation. Treat the results as estimates; the cited primary papers give the inputs.
  5. RCPAQAP Chemical Pathology. QAP Commutability & Matrix Effects. 2013.
  6. EPTIS (BAM) database entry 182373: RCPAQAP General Serum Chemistry (lyophilised serum with matching diluent; NATA ISO/IEC 17043).
  7. RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/
  8. EFLM Biological Variation Database (Aarsand AK, Fernandez-Calle P, Webster C, et al., EFLM Working Group on Biological Variation). Meta-analysis medians and 95% CIs read from the public API https://biologicalvariation.eu/api/meta_calculations; analyte list from /api/analytes and individual studies from /api/bv_specifications; retrieved 2026-09-29.