Free thyroxine (FT4) EQA results: interpretation and investigation guidance
The RCPAQAP limits for free thyroxine (FT4), known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the free thyroxine (FT4) monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Check the time the reconstituted vial stood at room temperature.
Limits, units and performance goals for free thyroxine (FT4)
| RCPAQAP APS, General Serum Chemistry (GSC) | ±1.5 pmol/L up to 12 pmol/L, ±12% above |
|---|---|
| Units | pmol/L. 1 ng/dL = 12.871 pmol/L |
| Standardisation | Partly standardised |
| Expected error mode | Proportional |
| Biological variation | Within-subject CV 4.8% and between-subject CV 8.01%, from the EFLM database (meta-analysis, 5 studies). |
| Desirable performance | Imprecision 2.4%, bias 2.33%, total allowable error 6.29%, derived from biological variation (Fraser). |
The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.
Known sources of error in free thyroxine (FT4) EQA results
In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.
Method-group target and commutability of the EQA material
Immunoassays for this measurand are not harmonised: calibration traceability, antibodies and the handling of processed (lyophilised, pooled or spiked) material differ between methods, so differences between method groups in EQA material may not exist in patient samples.[1, 2, 3, 4]
Binding-protein behaviour of processed material
A direct free T4 assay measures the free fraction left by the sample's binding proteins; lyophilised, pooled or T4-spiked material can have altered binding capacity, so methods that disturb the equilibrium differently diverge on that material.[3, 5, 1]
Recalibration or calibrator handling
A reconstitution volume error, an evaporated or out-of-date opened or reconstituted calibrator, or a calibrator from the wrong lot shifts every result after that calibration on that analyser.
New reagent lot
Immunoassay reagent lots differ in antibody and conjugate behaviour.[6]
Chemiluminescent signal or wash fault in a competitive assay
In a competitive assay the signal falls as analyte concentration rises.[7]
Low-end imprecision
Relative imprecision rises as concentration falls towards the lower limit of the measuring interval, so a small absolute difference is a large percentage on a low sample.[8]
Investigation checks in suggested order
- Check the time the reconstituted vial stood at room temperature.
- Check units: ng/dL x 12.9 = pmol/L; do not treat FT4 as dilution-linear.
- Compare with the laboratory's own FT4 method group; FT4 is not harmonised between methods.
- Review the FT4 calibration dates and reagent lot changes against the analysis date; compare IQC before and after.
EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs
The full free thyroxine (FT4) monograph in EQAWarden
This page is a summary. The monograph EQAWarden ships with holds 30 entries on what can go wrong with free thyroxine (FT4) in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each free thyroxine (FT4) result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about free thyroxine (FT4) opens it.
Questions about free thyroxine (FT4) EQA
What is the RCPAQAP analytical performance specification for free thyroxine (FT4)?
In the General Serum Chemistry (GSC) programme it is ±1.5 pmol/L up to 12 pmol/L, ±12% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.
What is a common cause of a free thyroxine (FT4) EQA result outside the APS?
Method-group target and commutability of the EQA material. Immunoassays for this measurand are not harmonised: calibration traceability, antibodies and the handling of processed (lyophilised, pooled or spiked) material differ between methods, so differences between method groups in EQA material may not exist in patient samples.
What should be checked first when a free thyroxine (FT4) EQA result is outside the APS?
Check the time the reconstituted vial stood at room temperature.
How are free thyroxine (FT4) units converted?
1 ng/dL = 12.871 pmol/L.
Where every fact on this page came from.
- Miller WG, Jones GRD, Horowitz GL, Weykamp C. Proficiency testing/external quality assessment: current challenges and future directions. Clin Chem 2011;57:1670-80. doi:10.1373/clinchem.2011.168641
- Badrick T, Miller WG, Panteghini M, Delatour V, Berghall H, MacKenzie F, Jones G. Interpreting EQA - understanding why commutability of materials matters. Clin Chem 2022;68:494-500. doi:10.1093/clinchem/hvac002
- Thienpont LM, Van Uytfanghe K, Beastall G, et al. Report of the IFCC Working Group for Standardization of Thyroid Function Tests; part 2: free thyroxine and free triiodothyronine. Clin Chem 2010;56:912-920.
- De Grande LAC, Van Uytfanghe K, Reynders D, et al. Standardization of free thyroxine measurements allows the adoption of a more uniform reference interval. Clin Chem 2017;63:1642-52. doi:10.1373/clinchem.2017.274407
- Miller WG, Schimmel H, Rej R, et al. IFCC Working Group recommendations for assessing commutability Part 1: general experimental design. Clin Chem 2018;64:447-54. doi:10.1373/clinchem.2017.277525
- Miller WG, Erek A, Cunningham TD, Oladipo O, Scott MG, Johnson RE. Commutability limitations influence quality control results with different reagent lots. Clin Chem 2011;57:76-83. doi:10.1373/clinchem.2010.148106
- Wild D (ed.). The Immunoassay Handbook: Theory and Applications of Enzyme Immunoassay, Electrochemiluminescence and Related Techniques. 4th ed. Oxford: Elsevier; 2013.
- RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/
- Li D, Radulescu A, Shrestha RT, et al. Association of biotin ingestion with performance of hormone and nonhormone assays in healthy adults. JAMA 2017;318:1150-1160.
- Colon PJ, Greene DN. Biotin interference in clinical immunoassays. J Appl Lab Med 2018;2:941-951.
- Kricka LJ. Human anti-animal antibody interferences in immunological assays. Clin Chem 1999;45:942-956.
- Tate J, Ward G. Interferences in immunoassay. Clin Biochem Rev 2004;25:105-120 (AACB journal). https://pmc.ncbi.nlm.nih.gov/articles/PMC1904417/
See how EQAWarden reviews a free thyroxine (FT4) result
Twenty minutes on a video call, on a fictional demo year: the score, the findings, the ranked causes and the signed review.