EQAWarden knowledge base · Digoxin

Digoxin EQA results: interpretation and investigation guidance

The RCPAQAP limits for digoxin, known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the digoxin monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Check reagent lot and pack onboard age, and IQC at the lowest level against the 0.2 µg/L APS limit.

At a glance

Limits, units and performance goals for digoxin

RCPAQAP APS, General Serum Chemistry (GSC)±0.2 µg/L up to 1.6 µg/L, ±10% above
Unitsµg/L. 1 ng/mL = 1 µg/L; 1 nmol/L = 0.78094 µg/L
StandardisationPartly standardised
Expected error modeMixed
Desirable performancetotal allowable error 10.0%, derived from biological variation (Fraser). Bias and total error are not calculated where the database holds no between-subject figure.

The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.

Known EQA pitfalls

Known sources of error in digoxin EQA results

In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.

Method-group offset at low and therapeutic concentrations

Digoxin immunoassays differ by method; a constant offset between methods is largest in relative terms at low targets, where the APS absolute limit (0.2 µg/L up to 1.6 µg/L) applies.[1, 2]

Recalibration or calibrator handling

A calibration sets the response for every later result on that analyser.

Glycoside metabolites or preservatives in EQA material

Digoxin immunoassays vary in cross-reactivity with digoxin metabolites and other cardiac glycosides.[3, 1, 4]

Sample pre-dilution on the analyser (sample aspiration, diluent volume)

Where the analyser pre-dilutes the sample before the assay, an inaccurate sample aspiration or diluent delivery changes the dilution factor for every pre-diluted assay on that analyser.

Spiked, lyophilised EQA material and commutability

RCPAQAP General Serum Chemistry is lyophilised pooled material with analyte additions that RCPAQAP says must not be assumed commutable.[5, 6, 7, 8, 9]

Measuring interval limits and dilution

A sample near or beyond the measuring interval depends on the automatic or manual dilution applied and on how a result outside the interval is reported.

Unit or transcription error in the submitted result

µg/L, ng/mL and nmol/L are all in use; nmol/L versus µg/L gives a ratio of 1.28 (or 0.78).[10, 11]

Investigation checks

Investigation checks in suggested order

  1. Check reagent lot and pack onboard age, and IQC at the lowest level against the 0.2 µg/L APS limit.
  2. Check the analyser HIL indices for the EQA sample.
  3. Confirm which analyser and assay reported digoxin and that the EQA method registration matches.
  4. Compare with the same-method peer group; regress the cycle on targets to separate an intercept (method offset, low-level effects) from a slope (calibration).
  5. Review the calibrator: reconstitution or opening date, storage, lot, and the last calibration dates.
  6. Confirm the submitted value and unit against the analyser or LIS record and the EQA provider's unit; test the result-to-target ratio against the unit factors.

Unit traps

  • ng/mL and µg/L are numerically equal.
  • Ratio about 1.28 (or 0.78): nmol/L versus µg/L (M = 780.94 g/mol).

EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs

In EQAWarden

The full digoxin monograph in EQAWarden

This page is a summary. The monograph EQAWarden ships with holds 32 entries on what can go wrong with digoxin in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each digoxin result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about digoxin opens it.

Plainly answered

Questions about digoxin EQA

What is the RCPAQAP analytical performance specification for digoxin?

In the General Serum Chemistry (GSC) programme it is ±0.2 µg/L up to 1.6 µg/L, ±10% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.

What is a common cause of a digoxin EQA result outside the APS?

Method-group offset at low and therapeutic concentrations. Digoxin immunoassays differ by method; a constant offset between methods is largest in relative terms at low targets, where the APS absolute limit (0.2 µg/L up to 1.6 µg/L) applies.

What should be checked first when a digoxin EQA result is outside the APS?

Check reagent lot and pack onboard age, and IQC at the lowest level against the 0.2 µg/L APS limit.

How are digoxin units converted?

1 ng/mL = 1 µg/L; 1 nmol/L = 0.78094 µg/L.

References

Where every fact on this page came from.

  1. Dasgupta A. Significant improvement in digoxin immunoassays over four decades: newer assays are less affected by interferences. Ther Drug Monit 2023;45:26-34. https://pubmed.ncbi.nlm.nih.gov/36624574/
  2. RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/
  3. Dasgupta A. Impact of interferences including metabolite crossreactivity on therapeutic drug monitoring results. Ther Drug Monit 2012;34:496-506. https://pubmed.ncbi.nlm.nih.gov/22846896/
  4. Devleeschouwer N, Libeer JC, Schiettecatte J, et al. Positive interference of merthiolate in the TDx digoxin assay in control samples used in the Belgian External Quality Assessment (EQA) Scheme. Eur J Clin Chem Clin Biochem 1994;32:935-7. https://pubmed.ncbi.nlm.nih.gov/7696443/
  5. RCPAQAP Chemical Pathology. QAP Commutability & Matrix Effects. 2013.
  6. Wilson JF, Tsanaclis LM, Perrett JE, et al. Performance of techniques for measurement of therapeutic drugs in serum: a comparison based on external quality assessment data. Ther Drug Monit 1992;14:98-106. https://pubmed.ncbi.nlm.nih.gov/1585409/
  7. Tsanaclis LM, Wilson JF. Comparison by external quality assessment of performance of analytical systems for measurement of therapeutic drugs in serum. Ther Drug Monit 1997;19:420-6. https://pubmed.ncbi.nlm.nih.gov/9263383/
  8. Robijns K, Boone NW, Kuypers AW, et al. A multilaboratory commutability evaluation of proficiency testing material for carbamazepine and valproic acid: a study within the framework of the Dutch Calibration 2000 project. Ther Drug Monit 2015;37:445-50. https://pubmed.ncbi.nlm.nih.gov/25525758/
  9. Miller WG, Jones GRD, Horowitz GL, Weykamp C. Proficiency testing/external quality assessment: current challenges and future directions. Clin Chem 2011;57:1670-80.
  10. Farrell CJ, Jones G, Shepherd S, Tran D, Badrick T. Mass or molar units? Australia's drug problem persists. Pathology 2023;55:865-7. https://pubmed.ncbi.nlm.nih.gov/37248117/
  11. IUPAC standard atomic weights (abridged, CIAAW) used to compute molar masses: H 1.008, C 12.011, N 14.007, O 15.999, Na 22.990, Mg 24.305, P 30.974, Cl 35.45, K 39.098, Ca 40.078, Fe 55.845, I 126.904, Li 6.94. https://www.ciaaw.org/atomic-weights.htm
  12. RCPAQAP. Chemical Pathology Analytical Performance Specifications, December 2022 (PDF). https://rcpaqap.com.au/wp-content/uploads/2022/12/RCPA_302340_APS-Chemical-Pathology-Update-Dec-2022_FA.pdf

See how EQAWarden reviews a digoxin result

Twenty minutes on a video call, on a fictional demo year: the score, the findings, the ranked causes and the signed review.