C-reactive protein (CRP) EQA results: interpretation and investigation guidance
The RCPAQAP limits for C-reactive protein (CRP), known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the C-reactive protein (CRP) monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Check units (mg/L versus mg/dL, ratio 10) and transcription.
Limits, units and performance goals for C-reactive protein (CRP)
| RCPAQAP APS, C Reactive Protein | ±0.8 mg/L up to 4 mg/L, ±20% above |
|---|---|
| Units | mg/L. 1 mg/dL = 10 mg/L |
| Standardisation | Harmonised |
| Expected error mode | Mixed |
| Biological variation | Within-subject CV 34.68% and between-subject CV 87.08%, from the EFLM database (meta-analysis, 6 studies). |
| Desirable performance | Imprecision 17.34%, bias 23.43%, total allowable error 52.04%, derived from biological variation (Fraser). |
The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.
Known sources of error in C-reactive protein (CRP) EQA results
In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.
Standard and high-sensitivity CRP methods form separate groups
CRP methods differ in antibody, particle, calibrator and measuring range, and standard and high-sensitivity applications often run side by side; each forms its own method group.
RCPAQAP CRP material is not whole human serum
The RCPAQAP CRP programme material is described as an aqueous buffered protein matrix and is not assumed commutable, so method groups may separate on it in a way patient samples would not.[1, 2]
Reagent lot change at low CRP
A change of reagent lot in a particle-enhanced CRP immunoassay can shift results; a small absolute shift is a large relative change at low concentrations, so low samples can move while mid and high samples barely move.
Low-CRP samples near the quantitation limit
Imprecision rises as concentration approaches a method's limit of quantitation, and standard CRP methods have a higher lower limit than high-sensitivity methods.
Dilution of high samples on high-sensitivity methods
High-sensitivity CRP methods have a narrow measuring range, so a high EQA sample is reported after dilution, which adds its own error.
Antigen excess (hook)
Immunoturbidimetric and immunonephelometric CRP methods can under-read at antigen excess, but EQA materials are normally made at concentrations well below where this is expected.[3]
Rheumatoid factor and lipaemia at low CRP
Lipaemia adds turbidity and light scatter in turbidimetric and nephelometric methods, and rheumatoid factor can cross-link antibody-coated particles; both matter most at low CRP.
Reconstitution volume error or evaporation of the EQA vial
Too much or too little diluent at reconstitution, or water loss from an open or poorly stoppered vial, scales every dilution-linear analyte in that vial by the same percentage.
Correlation factor, unit conversion or LIS mapping in the result path
A correlation slope or intercept saved in the analyser's assay settings, or a conversion in middleware or the LIS, is applied to every result for that assay, EQA included; RCPAQAP expects EQA results without such factors.[4]
Investigation checks in suggested order
- Check units (mg/L versus mg/dL, ratio 10) and transcription.
- Check the calibrator opening date against the calibration in force.
- Check for automatic repeat or dilution flags on high results from a high-sensitivity assay.
- Confirm which CRP method (standard or high-sensitivity) produced the result and compare with that peer group.
- For low samples, check reagent lot changes and the method's imprecision at low levels.
EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs
The full C-reactive protein (CRP) monograph in EQAWarden
This page is a summary. The monograph EQAWarden ships with holds 29 entries on what can go wrong with C-reactive protein (CRP) in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each C-reactive protein (CRP) result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about C-reactive protein (CRP) opens it.
Questions about C-reactive protein (CRP) EQA
What is the RCPAQAP analytical performance specification for C-reactive protein (CRP)?
In the C Reactive Protein programme it is ±0.8 mg/L up to 4 mg/L, ±20% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.
What is a common cause of a C-reactive protein (CRP) EQA result outside the APS?
Standard and high-sensitivity CRP methods form separate groups. CRP methods differ in antibody, particle, calibrator and measuring range, and standard and high-sensitivity applications often run side by side; each forms its own method group.
What should be checked first when a C-reactive protein (CRP) EQA result is outside the APS?
Check units (mg/L versus mg/dL, ratio 10) and transcription.
How are C-reactive protein (CRP) units converted?
1 mg/dL = 10 mg/L.
Where every fact on this page came from.
- RCPAQAP. Products - Chemical Pathology listing pages (Wayback captures 2025-01-13 to 2026-03-09).
- RCPAQAP Chemical Pathology. QAP Commutability & Matrix Effects. 2013.
- Jassam N, Jones CM, Briscoe T, Horner JH. The hook effect: a need for constant vigilance. Ann Clin Biochem 2006;43:314-317.
- myQAP help. Post Analytical Correction Factors (Wayback 2026-01-13).
- RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/