EQAWarden knowledge base · Cortisol

Cortisol EQA results: interpretation and investigation guidance

The RCPAQAP limits for cortisol, known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the cortisol monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: For urine cortisol, confirm direct or extracted procedure and the collection-volume calculation.

At a glance

Limits, units and performance goals for cortisol

RCPAQAP APS, General Serum Chemistry (GSC)±15 nmol/L up to 100 nmol/L, ±15% above
RCPAQAP APS, Endocrine: Special±15 nmol/L up to 100 nmol/L, ±15% above. Endocrine basis per C1 3.3. The value is verified; the programme section it sits in is inferred from the programme menus.
Unitsnmol/L. 1 µg/dL = 27.59 nmol/L
StandardisationPartly standardised
Expected error modeProportional
Biological variationWithin-subject CV 16.15% and between-subject CV 21.51%, from the EFLM database (meta-analysis, 4 studies).
Desirable performanceImprecision 8.07%, bias 6.72%, total allowable error 20.05%, derived from biological variation (Fraser).

The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.

Known EQA pitfalls

Known sources of error in cortisol EQA results

In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.

Cross-reacting steroids in EQA material

Cortisol immunoassay antibodies can cross-react with prednisolone, 11-deoxycortisol and other structurally related steroids, which LC-MS/MS separates; a material containing such steroids reads high in immunoassay groups compared with LC-MS/MS.[1]

Method-group target and commutability of the EQA material

Immunoassays for this measurand are not harmonised: calibration traceability, antibodies and the handling of processed (lyophilised, pooled or spiked) material differ between methods, so differences between method groups in EQA material may not exist in patient samples.[2, 3, 1]

New reagent lot and its master curve

Many immunoassays supply a lot-specific master curve that a short calibration adjusts; a new lot needs a new calibration, and a lot-to-lot difference that the calibration does not remove moves every result on every analyser using that lot.[4]

Chemiluminescent signal or wash fault in a competitive assay

In a competitive immunoassay the signal falls as analyte rises.[5]

Heterophile or human anti-animal antibodies

Immunoassays can be affected by heterophile or human anti-animal antibodies, giving falsely high or low results.[6, 7]

Investigation checks

Investigation checks in suggested order

  1. For urine cortisol, confirm direct or extracted procedure and the collection-volume calculation.
  2. Check units: µg/dL x 27.59 = nmol/L.
  3. Compare with the immunoassay peer group of the method in use and note the LC-MS/MS group; a sample-specific immunoassay excess suggests cross-reacting steroids.
  4. Review reagent lot changes and IQC; check other analytes calibrated from the same calibrator on the same analyser.

Unit traps

  • Ratio about 27.6 = µg/dL versus nmol/L; urine results need the collection volume (ug/24 h x 2.76 = nmol/24 h).

EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs

In EQAWarden

The full cortisol monograph in EQAWarden

This page is a summary. The monograph EQAWarden ships with holds 25 entries on what can go wrong with cortisol in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each cortisol result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about cortisol opens it.

Plainly answered

Questions about cortisol EQA

What is the RCPAQAP analytical performance specification for cortisol?

In the General Serum Chemistry (GSC) programme it is ±15 nmol/L up to 100 nmol/L, ±15% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.

What is a common cause of a cortisol EQA result outside the APS?

Cross-reacting steroids in EQA material. Cortisol immunoassay antibodies can cross-react with prednisolone, 11-deoxycortisol and other structurally related steroids, which LC-MS/MS separates; a material containing such steroids reads high in immunoassay groups compared with LC-MS/MS.

What should be checked first when a cortisol EQA result is outside the APS?

For urine cortisol, confirm direct or extracted procedure and the collection-volume calculation.

How are cortisol units converted?

1 µg/dL = 27.59 nmol/L.

References

Where every fact on this page came from.

  1. JCTLM Database of higher-order reference materials, measurement methods/procedures and services (queried 2026-09-28 via the public search API https://www.jctlmdb.org/backend/api/jctlm/searchRmp|searchRm|searchRms). https://www.jctlmdb.org/
  2. Miller WG, Jones GRD, Horowitz GL, Weykamp C. Proficiency testing/external quality assessment: current challenges and future directions. Clin Chem 2011;57:1670-80. doi:10.1373/clinchem.2011.168641
  3. Badrick T, Miller WG, Panteghini M, Delatour V, Berghall H, MacKenzie F, Jones G. Interpreting EQA - understanding why commutability of materials matters. Clin Chem 2022;68:494-500. doi:10.1093/clinchem/hvac002
  4. Miller WG, Erek A, Cunningham TD, Oladipo O, Scott MG, Johnson RE. Commutability limitations influence quality control results with different reagent lots. Clin Chem 2011;57:76-83. doi:10.1373/clinchem.2010.148106
  5. Wild D (ed.). The Immunoassay Handbook: Theory and Applications of Enzyme Immunoassay, Electrochemiluminescence and Related Techniques. 4th ed. Oxford: Elsevier; 2013.
  6. Kricka LJ. Human anti-animal antibody interferences in immunological assays. Clin Chem 1999;45:942-956.
  7. Tate J, Ward G. Interferences in immunoassay. Clin Biochem Rev 2004;25:105-120 (AACB journal). https://pmc.ncbi.nlm.nih.gov/articles/PMC1904417/
  8. RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/