Creatine kinase (CK) EQA results: interpretation and investigation guidance
The RCPAQAP limits for creatine kinase (CK), known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the creatine kinase (CK) monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Record the time from reconstitution to CK measurement (activity falls on standing).
Limits, units and performance goals for creatine kinase (CK)
| RCPAQAP APS, General Serum Chemistry (GSC) | ±15 U/L up to 125 U/L, ±12% above |
|---|---|
| Units | U/L. 1 ukat/L = 60 U/L |
| Standardisation | Traceable to a reference method |
| Expected error mode | Proportional |
| Biological variation | Within-subject CV 14.12% and between-subject CV 32.92%, from the EFLM database (meta-analysis, 3 studies). |
| Desirable performance | Imprecision 7.06%, bias 8.96%, total allowable error 20.6%, derived from biological variation (Fraser). |
The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.
Known sources of error in creatine kinase (CK) EQA results
In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.
CK activity loss after reconstitution
CK is thiol-dependent and loses activity by oxidation after reconstitution; light and freezing also lower activity in some materials; the N-acetylcysteine in the reagent re-activates only partly[1, 2].
Reaction temperature away from 37 C
Catalytic activity rises with temperature, so every enzyme read in a cuvette that is warmer or cooler than 37 C is proportionally high or low.[3, 4]
Haemolysis in a low-activity sample
Red cells release adenylate kinase and other components that add to the coupled CK reaction, so haemolysis raises CK; the relative effect is largest at low CK.
EQA vial reconstitution volume or evaporation
Under- or over-dilution when the lyophilised vial is reconstituted, or water loss from an open vial or sample cup, changes every solute and enzyme activity in that vial by the same proportion.[5]
Investigation checks in suggested order
- Record the time from reconstitution to CK measurement (activity falls on standing).
- Look at all enzymes of the same survey: CK moves most with a temperature error.
- Check the sample description for haemolysis at low CK.
- Compute the percentage deviation of other analytes in the same vial.
- Check units (ratio 60 for ukat/L).
Unit traps
- Ratio 60 (or 0.0167) = ukat/L vs U/L.
- CK (activity, U/L) and CK-MB (mass, µg/L) must not be crossed in the mapping.
EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs
The full creatine kinase (CK) monograph in EQAWarden
This page is a summary. The monograph EQAWarden ships with holds 18 entries on what can go wrong with creatine kinase (CK) in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each creatine kinase (CK) result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about creatine kinase (CK) opens it.
Questions about creatine kinase (CK) EQA
What is the RCPAQAP analytical performance specification for creatine kinase (CK)?
In the General Serum Chemistry (GSC) programme it is ±15 U/L up to 125 U/L, ±12% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.
What is a common cause of a creatine kinase (CK) EQA result outside the APS?
CK activity loss after reconstitution. CK is thiol-dependent and loses activity by oxidation after reconstitution; light and freezing also lower activity in some materials; the N-acetylcysteine in the reagent re-activates only partly.
What should be checked first when a creatine kinase (CK) EQA result is outside the APS?
Record the time from reconstitution to CK measurement (activity falls on standing).
How are creatine kinase (CK) units converted?
1 ukat/L = 60 U/L.
Where every fact on this page came from.
- Perry B, Doumas B, Jendrzejczak B. Effect of light and temperature on the stability of creatine kinase in human sera and controls. Clin Chem 1979;25:625-628.
- Creatine kinase activity decrease with short-term freezing (journal article listing; about 2.6%, 5.9% and 8.3% median loss at 24 h, 2 weeks and 4 weeks at -30 C). Authors and journal not confirmed.
- Schumann G, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 C. Part 2. Creatine kinase. Clin Chem Lab Med 2002;40.
- Derived arithmetic (not a publication): per-degree temperature coefficients near 37 C computed as (F)^(1/7)-1 from published 37/30 C conversion factors, or from activation energies with the Arrhenius equation. Treat the results as estimates; the cited primary papers give the inputs.
- EPTIS (BAM) database entry 182373: RCPAQAP General Serum Chemistry (lyophilised serum with matching diluent; NATA ISO/IEC 17043).
- RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/
- EFLM Biological Variation Database (Aarsand AK, Fernandez-Calle P, Webster C, et al., EFLM Working Group on Biological Variation). Meta-analysis medians and 95% CIs read from the public API https://biologicalvariation.eu/api/meta_calculations; analyte list from /api/analytes and individual studies from /api/bv_specifications; retrieved 2026-09-29.
- Fraser CG. Biological Variation: From Principles to Practice. Washington DC: AACC Press; 2001.
- Sandberg S, Fraser CG, Horvath AR, et al. Defining analytical performance specifications: Consensus Statement from the 1st Strategic Conference of the EFLM. Clin Chem Lab Med 2015;53:833-5.