EQAWarden knowledge base · Cholesterol

Cholesterol EQA results: interpretation and investigation guidance

The RCPAQAP limits for cholesterol, known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the cholesterol monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Check transcription and units (mmol/L versus mg/dL, ratio about 38.7) against the instrument printout.

At a glance

Limits, units and performance goals for cholesterol

RCPAQAP APS, General Serum Chemistry (GSC)±0.3 mmol/L up to 5 mmol/L, ±6% above
Unitsmmol/L. 1 mg/dL = 0.025862 mmol/L
StandardisationTraceable to a reference method
Expected error modeProportional
Biological variationWithin-subject CV 5.26% and between-subject CV 15.51%, from the EFLM database (meta-analysis, 36 studies).
Desirable performanceImprecision 2.63%, bias 4.09%, total allowable error 8.43%, derived from biological variation (Fraser).

The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.

Known EQA pitfalls

Known sources of error in cholesterol EQA results

In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.

Calibrator shared by several assays

Where one calibrator material serves several assays on an analyser, a reconstitution or volume error, an aged or wrongly stored vial, or a lot value entry error moves every assay calibrated from it by a similar percentage, at staggered dates when the assays are recalibrated at different times.

Peroxidase (Trinder) end-point interference from reductants

Cholesterol is measured with a peroxidase-coupled colour end point; reductants such as ascorbate, N-acetylcysteine and metamizole consume peroxide or the dye before it is read and depress the result.[1, 2]

Bilirubin in icteric EQA material

Bilirubin interferes with peroxidase-coupled cholesterol methods, usually lowering the result; the size and even the direction differ between methods[2].

Non-commutable lyophilised material

Processed, lyophilised lipid material can react differently from native serum in enzymatic methods, creating method-group differences that do not exist in patient samples; RCPAQAP lipid material is lyophilised serum without a public commutability statement.[3, 4, 5]

Reagent carry-over from a preceding test

Carry-over of a preceding reagent into the cholesterol reaction has been reported on automated analysers (intermittently low cholesterol after creatine kinase reagent); the effect depends on test order[6].

Reconstitution volume error or evaporation of the EQA vial

Too much or too little diluent at reconstitution, or water loss from an open or poorly stoppered vial, scales every dilution-linear analyte in that vial by the same percentage.

Correlation factor, unit conversion or LIS mapping in the result path

A correlation slope or intercept saved in the analyser's test settings, or a conversion in the middleware or LIS, is applied to every result for that assay, EQA included; RCPAQAP expects EQA results without such factors.[7]

Investigation checks

Investigation checks in suggested order

  1. Check transcription and units (mmol/L versus mg/dL, ratio about 38.7) against the instrument printout.
  2. Check H and I indices and the bilirubin target of the deviating sample.
  3. Compare with the same-method peer group before the all-method or reference target.
  4. Look at the other assays calibrated from the same calibrator on the same analyser and at the other peroxidase end-point analytes of the same sample for a shared shift.
  5. Check for a correlation factor or a unit conversion in the analyser's test settings, the middleware or the LIS.

Unit traps

  • Ratio about 38.7 (or 0.0259) = mg/dL versus mmol/L confusion.

EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs

In EQAWarden

The full cholesterol monograph in EQAWarden

This page is a summary. The monograph EQAWarden ships with holds 25 entries on what can go wrong with cholesterol in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each cholesterol result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about cholesterol opens it.

Plainly answered

Questions about cholesterol EQA

What is the RCPAQAP analytical performance specification for cholesterol?

In the General Serum Chemistry (GSC) programme it is ±0.3 mmol/L up to 5 mmol/L, ±6% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.

What is a common cause of a cholesterol EQA result outside the APS?

Calibrator shared by several assays. Where one calibrator material serves several assays on an analyser, a reconstitution or volume error, an aged or wrongly stored vial, or a lot value entry error moves every assay calibrated from it by a similar percentage, at staggered dates when the assays are recalibrated at different times.

What should be checked first when a cholesterol EQA result is outside the APS?

Check transcription and units (mmol/L versus mg/dL, ratio about 38.7) against the instrument printout.

How are cholesterol units converted?

1 mg/dL = 0.025862 mmol/L.

References

Where every fact on this page came from.

  1. Martinello F, Luiz da Silva E. Mechanism of ascorbic acid interference in biochemical tests that use peroxide and peroxidase to generate chromophore. Clin Chim Acta 2006;373:108-116.
  2. Spain MA, Wu AH. Bilirubin interference with determination of uric acid, cholesterol, and triglycerides in commercial peroxidase-coupled assays, and the effect of ferrocyanide. Clin Chem 1986;32:518-521.
  3. Miller WG, Jones GRD, Horowitz GL, Weykamp C. Proficiency testing/external quality assessment: current challenges and future directions. Clin Chem 2011;57:1670-80.
  4. RCPAQAP Chemical Pathology. QAP Commutability & Matrix Effects. 2013.
  5. EPTIS (BAM) database entry 183532: RCPAQAP Special Lipids (lyophilised serum).
  6. Kyle PB. Beware of carryover in modern chemistry analyzers. Clin Chem Lab Med 2010;48:519-521.
  7. myQAP help. Post Analytical Correction Factors (Wayback 2026-01-13).
  8. RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/