EQAWarden knowledge base · Bilirubin (total)

Bilirubin (total) EQA results: interpretation and investigation guidance

The RCPAQAP limits for bilirubin (total), known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the bilirubin (total) monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Ask about light protection and time from reconstitution to analysis; compare with conjugated bilirubin in the same vial.

At a glance

Limits, units and performance goals for bilirubin (total)

RCPAQAP APS, General Serum Chemistry (GSC)±3 µmol/L up to 25 µmol/L, ±12% above
Unitsµmol/L. 1 mg/dL = 17.104 µmol/L
StandardisationPartly standardised
Expected error modeProportional
Biological variationWithin-subject CV 20.18% and between-subject CV 24.9%, from the EFLM database (meta-analysis, 8 studies).
Desirable performanceImprecision 10.09%, bias 8.01%, total allowable error 24.66%, derived from biological variation (Fraser).

The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.

Known EQA pitfalls

Known sources of error in bilirubin (total) EQA results

In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.

Vanadate versus diazo method group

Vanadate oxidation and diazo methods differ in chemistry and calibrator; published comparisons show vanadate reading about 17-20% above diazo or RMP values in neonatal samples, with matrix-dependent behaviour[1, 2, 3].

Light exposure of the reconstituted sample

Bilirubin photo-oxidises and isomerises in room and window light; photoproducts react differently with different chemistries[4, 5].

Bilirubin calibrator handling or a change of calibrator

Bilirubin in a reconstituted calibrator degrades with light and time, so an aged or light-exposed calibrator vial changes the calibration factor; a change of calibrator product can appear as a step.[4]

Freeze-thaw of liquid serum material

Total bilirubin was unstable beyond one freeze-thaw cycle in serum[6].

EQA vial reconstitution volume or evaporation

Under- or over-dilution when the lyophilised vial is reconstituted, or water loss from an open vial or sample cup, changes every solute and enzyme activity in that vial by the same proportion.[7]

Investigation checks

Investigation checks in suggested order

  1. Ask about light protection and time from reconstitution to analysis; compare with conjugated bilirubin in the same vial.
  2. At low concentrations assess the deviation in µmol/L against low-level IQC imprecision before calling it a bias.
  3. Check the sample description for haemolysis or lipaemia.
  4. Check units (µmol/L vs mg/dL, ratio 17.1) and any unit factor configured on the analyser.
  5. Identify the chemistry (vanadate or diazo) and compare with that method group; expect vanadate and diazo groups to separate.
  6. Review the calibrator used, its time from reconstitution to calibration and its light protection, and the calibration dates.

Unit traps

  • Conjugated and total bilirubin swapped in the result mapping gives a large sample-specific negative (or positive) deviation.

EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs

In EQAWarden

The full bilirubin (total) monograph in EQAWarden

This page is a summary. The monograph EQAWarden ships with holds 23 entries on what can go wrong with bilirubin (total) in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each bilirubin (total) result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about bilirubin (total) opens it.

Plainly answered

Questions about bilirubin (total) EQA

What is the RCPAQAP analytical performance specification for bilirubin (total)?

In the General Serum Chemistry (GSC) programme it is ±3 µmol/L up to 25 µmol/L, ±12% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.

What is a common cause of a bilirubin (total) EQA result outside the APS?

Vanadate versus diazo method group. Vanadate oxidation and diazo methods differ in chemistry and calibrator; published comparisons show vanadate reading about 17-20% above diazo or RMP values in neonatal samples, with matrix-dependent behaviour.

What should be checked first when a bilirubin (total) EQA result is outside the APS?

Ask about light protection and time from reconstitution to analysis; compare with conjugated bilirubin in the same vial.

How are bilirubin (total) units converted?

1 mg/dL = 17.104 µmol/L.

References

Where every fact on this page came from.

  1. Westenberg LEH, Lantman MVS, van Berkel M, et al. Evaluation of total bilirubin in a national quality program. Clin Chem 2025;72:183-191.
  2. Ryan-Beswick F, Kyle C, Parke L, et al. Comparability of total bilirubin methods for neonatal samples across 6 NHS trusts within North West England. Ann Clin Biochem 2026 (online ahead of print).
  3. Thomas DH, Warner JV, Jones GRD, et al. Total bilirubin assay differences may cause inconsistent treatment decisions in neonatal hyperbilirubinaemia. Clin Chem Lab Med 2022;60:1736-1744.
  4. Sofronescu AG, Loebs T, Zhu Y. Effects of temperature and light on the stability of bilirubin in plasma samples. Clin Chim Acta 2012;413:463-466.
  5. Ihara H, Nakamura H, Aoki Y, Aoki T, Yoshida M. In vitro effects of light on serum bilirubin subfractions measured by high-performance liquid chromatography: comparison with four routine methods. Clin Chem 1992;38:2124-2129.
  6. Cuhadar S, Koseoglu M, Atay A, Dirican A. The effect of storage time and freeze-thaw cycles on the stability of serum samples. Biochem Med (Zagreb) 2013;23:70-77.
  7. EPTIS (BAM) database entry 182373: RCPAQAP General Serum Chemistry (lyophilised serum with matching diluent; NATA ISO/IEC 17043).
  8. RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/
  9. EFLM Biological Variation Database (Aarsand AK, Fernandez-Calle P, Webster C, et al., EFLM Working Group on Biological Variation). Meta-analysis medians and 95% CIs read from the public API https://biologicalvariation.eu/api/meta_calculations; analyte list from /api/analytes and individual studies from /api/bv_specifications; retrieved 2026-09-29.
  10. Fraser CG. Biological Variation: From Principles to Practice. Washington DC: AACC Press; 2001.
  11. Sandberg S, Fraser CG, Horvath AR, et al. Defining analytical performance specifications: Consensus Statement from the 1st Strategic Conference of the EFLM. Clin Chem Lab Med 2015;53:833-5.