EQAWarden knowledge base · Albumin

Albumin EQA results: interpretation and investigation guidance

The RCPAQAP limits for albumin, known reasons a result can depart from its target in external quality assessment (EQA) and the checks to make. A summary of the albumin monograph EQAWarden ships with; the full monograph is in EQAWarden. First check: Check whether the deviation is larger at the low-albumin sample (BCG globulin effect) or constant in percentage (calibration).

At a glance

Limits, units and performance goals for albumin

RCPAQAP APS, General Serum Chemistry (GSC)±2 g/L up to 33 g/L, ±6% above
Unitsg/L. 1 g/dL = 10 g/L
StandardisationPartly standardised
Expected error modeMixed
Biological variationWithin-subject CV 2.5% and between-subject CV 4.32%, from the EFLM database (meta-analysis, 16 studies).
Desirable performanceImprecision 1.25%, bias 1.25%, total allowable error 3.31%, derived from biological variation (Fraser).

The methods, their calibration and what is particular to each platform are in the full monograph in EQAWarden. Your own instructions for use come first.

Known EQA pitfalls

Known sources of error in albumin EQA results

In general terms. How each one shows in the results, which way it moves them and what to check for it are in the full monograph in EQAWarden.

BCG versus BCP method group

BCG also binds some globulins (alpha1, alpha2 and acute-phase proteins), overestimating albumin by about 3.5 g/L against electrophoresis and up to about 6 g/L in nephrotic sera, while BCP agrees within about 1 g/L[1, 2, 3].

Shared multi-analyte calibrator handling or lot value

Where one calibrator serves several assays, a reconstitution volume error, an aged vial or a mis-entered lot value shifts all the assays calibrated from it together, in proportion.

EQA vial reconstitution volume or evaporation

Under- or over-dilution when the lyophilised vial is reconstituted, or water loss from an open vial or sample cup, changes every solute and enzyme activity in that vial by the same proportion.[4]

Unit and conversion factor configuration

An inconsistent unit or conversion factor (g/dL x 10 = g/L) between analyser, middleware and LIS gives a constant ratio.

Investigation checks

Investigation checks in suggested order

  1. Check whether the deviation is larger at the low-albumin sample (BCG globulin effect) or constant in percentage (calibration).
  2. Compute the percentage deviation of the other analytes in the same sample for a reconstitution or predilution signature.
  3. Check the sample description for haemolysis or lipaemia.
  4. Check units (g/L vs g/dL) and the configured conversion factor.
  5. Identify the dye (BCG or BCP) and confirm the RCPAQAP method code; compare with that dye's peer group.
  6. Review the calibrator's reconstitution time, lot value entry and calibration dates; compare the other assays of the same survey that share the calibrator.

Unit traps

  • A urine albumin (mg/L) result mapped to serum albumin gives a ratio near 1000.

EQAWarden applies its documented rules to each survey and ranks potential causes by supporting evidence; the reviewer reaches the conclusion. All analytes · The 59 rules · The whole knowledge base · How an EQA investigation runs

In EQAWarden

The full albumin monograph in EQAWarden

This page is a summary. The monograph EQAWarden ships with holds 23 entries on what can go wrong with albumin in EQA, general and by platform: for each, how it shows in the results, which way it moves them, how sure the evidence is and what to check. It also holds the method and calibration detail and what is particular to Siemens Atellica, Roche cobas, and Abbott Alinity and ARCHITECT. EQAWarden uses it when assessing each albumin result and ranks potential causes by their supporting evidence; the reviewer reaches the conclusion. In EQAWarden the whole monograph is under Training, Monographs, shown for the platforms of the laboratory's own analysers with a switch to every platform, and each finding about albumin opens it.

Plainly answered

Questions about albumin EQA

What is the RCPAQAP analytical performance specification for albumin?

In the General Serum Chemistry (GSC) programme it is ±2 g/L up to 33 g/L, ±6% above, evaluated at the target. RCPAQAP can change its specifications, so confirm the current table at each enrolment year.

What is a common cause of an albumin EQA result outside the APS?

BCG versus BCP method group. BCG also binds some globulins (alpha1, alpha2 and acute-phase proteins), overestimating albumin by about 3.5 g/L against electrophoresis and up to about 6 g/L in nephrotic sera, while BCP agrees within about 1 g/L.

What should be checked first when an albumin EQA result is outside the APS?

Check whether the deviation is larger at the low-albumin sample (BCG globulin effect) or constant in percentage (calibration).

How are albumin units converted?

1 g/dL = 10 g/L.

References

Where every fact on this page came from.

  1. Garcia Moreira V, Beridze Vaktangova N, Martinez Gago MD, et al. Overestimation of albumin measured by bromocresol green vs bromocresol purple method: influence of acute-phase globulins. Lab Med 2018;49:355-361.
  2. van de Logt AE, Rijpma SR, Vink CH, et al. The bias between different albumin assays may affect clinical decision-making. Kidney Int 2019;95:1514-1517.
  3. Ueno T, Hirayama S, Sugihara M, Miida T. The bromocresol green assay, but not the modified bromocresol purple assay, overestimates the serum albumin concentration in nephrotic syndrome through reaction with alpha2-macroglobulin. Ann Clin Biochem 2016;53:97-105.
  4. EPTIS (BAM) database entry 182373: RCPAQAP General Serum Chemistry (lyophilised serum with matching diluent; NATA ISO/IEC 17043).
  5. RCPAQAP. Chemical Pathology Analytical Performance Specifications (web page; dateModified 2025-03-05). https://rcpaqap.com.au/resources/chemical-pathology-analytical-performance-specifications/
  6. EFLM Biological Variation Database (Aarsand AK, Fernandez-Calle P, Webster C, et al., EFLM Working Group on Biological Variation). Meta-analysis medians and 95% CIs read from the public API https://biologicalvariation.eu/api/meta_calculations; analyte list from /api/analytes and individual studies from /api/bv_specifications; retrieved 2026-09-29.
  7. Fraser CG. Biological Variation: From Principles to Practice. Washington DC: AACC Press; 2001.
  8. Sandberg S, Fraser CG, Horvath AR, et al. Defining analytical performance specifications: Consensus Statement from the 1st Strategic Conference of the EFLM. Clin Chem Lab Med 2015;53:833-5.